Cpu Board Switch Settings; Table 2-1. Sw1 Dip Switch Settings - Thermo Scientific Oretronic III Operating & Service Manual

Table of Contents

Advertisement

POS
FUNCTION
1
Password protection
2
Speed sensor input
3
SYNC input
4
Belt direction
5
Factory test
6
Units selection
7
F5
8
Cold start
CPU Board Switch
Settings
Thermo Fisher Scientific
Enable password protection software.
Enable speed sensor input option.
Enable SYNC input on the slaves for
synchronization of transmitter pulses.
(Required when two detectors are within 30
ft of each other.)
Reversed
Metric
Enable F5 error alarm
Enable cold start on power up.
Table 2–1. SW1 DIP Switch Settings
Four switches on the CPU board must be set for power, line voltage, and
alarm output for the TMD. If RS-485 communication is supported, a
fifth switch must also be set.
Figure 2-4 shows the locations of the switches on the CPU board.
Table 2-2 and Table 2-3 describe the functions of the switch settings.
ON
Installing the TMD
OFF
Eliminate password.
Speed sensor disabled.
SYNC input disabled.
(Required setting on the
master.)
Normal
Normal operation.
(Required setting.)
English
Disables F5
Cold start disabled.
Installing the TMD
REC-4105 Rev L
2-13