Installing the TMD
Mounting
Location
Electrical
Specifications
Input Power
Requirements
2-2
Installing the TMD
REC-4105 Rev L
The coil support structure and coil(s) must be mounted properly for the
TMD to function correctly. Large metallic objects near the coil(s) will
cause a reduction in signal strength. To prevent loss in signal strength,
make sure that no metallic objects such as steel girders, deck plates, skirt
boards, return idlers, or metal building walls are within 4 ft (1.2 m) of
the receiver coil. (Carry idlers and conveyor stringers may be within
this distance.)
Other considerations for mounting the support structure and coil(s)
include: The coil support structure must be located so that the conveyor
will stop before tramp is discharged.
Metal deck plates or skirt boards located within 4 ft (1.2 m) either
side of the coil support structure must be removed or replaced with
non-metallic parts.
Metal chutes of feed points should be at least 8 ft (2.4 m) away from
the coil support structure.
Magnets must be at least 18 ft (5.5 m) away from the coil support
structure.
Corrugated metal covers located within 4 ft (1.2 m) either side of the
coil support structure must be removed or replaced with non-metallic
covers.
The control unit enclosure should be mounted in a vibration-free area
less than 10 ft (3 m) from the coil(s) and should be protected from
excessive heat, cold, or moisture.
See the Final Assembly drawings and the Installation Placement
drawings in Appendix D of this manual for additional installation
considerations.
115/230 VAC, +10%, –15%
50/60 Hz
Single phase, 30 VA, 0.25 A
Switch selectable: 115 VAC or 230 VAC
Fusing:
115 VAC (F1) - 3/8 A, 250 VAC, SB, "Type T"
230 VAC (F1) - 3/16 A, 250 VAC, SB, "Type T"