14 Insert pump piston
Pump piston
16 Insert the heat exchanger capillary into the
outlet of the primary pump head. Using a
torque key, set 3 Nm and close the hex screw
at the top of the outlet.
©
*G4220-90120*
*G4220-90120*
Agilent Technologies 2010
Part Number:
G4220-90120
Edition:
15 Insert both pump heads in the link plate and
make sure that the clips snap in that fix the
pump heads.
Link plate
Clips
17 Install the pump head assembly and install
flow connections as described in Replacing
the pump heads, discribed in the user Manual.
04/10
Printed in Germany
Agilent Technologies
Hewlett-Packard-Strasse 8
76337 Waldbronn