DFS150-485 2D Fixed Scanner
Chapter 1: Overview .............................................................. 1-1
Chapter 2: Interface .............................................................. 2-1
2.1 Input and Output ........................................................................... 2-2
2.2 RS485 ............................................................................................ 2-4
2.2.1 Features ............................................................................... 2-4
2.3 LED & Beeper ............................................................................... 2-6
Chapter 3: Power On ............................................................. 3-1
Chapter 4: Operation Mode ................................................. 4-1
4.1 Master Private Mode ..................................................................... 4-2
4.2 Master Modbus Mode (PLC Link) ................................................ 4-3
4.3 Slave Modbus Mode ..................................................................... 4-5
Chapter 5: Trigger .................................................................. 5-1
5.1 Continuous Trigger ...................................................................... 5-1
5.2 I/O Trigger ...................................................................................... 5-1
5.3 Command Trigger ......................................................................... 5-2
Appendix ................................................................................... A-1
Continuous mode reread delay ......................................................... A-4
Manual mode centering windows configuration .............................. A-5
Data Editing ........................................................................................ A-7
Manual mode centering windows configuration .............................. A-5
Data Editing ........................................................................................ A-7
Symbologies ....................................................................................... A-8
Field of View .................................................................................... A-16
Working Range ................................................................................ A-17
Modbus Exception Code ................................................................. A-18
DFS150-485 2D Fixed Scanner User Guide
User Guide

Table of Contents

V1.0

Advertisement

Table of Contents
loading

Summary of Contents for Delta DFS150-485

  • Page 1: Table Of Contents

    Manual mode centering windows configuration ......A-5 Data Editing ..................A-7 Manual mode centering windows configuration ......A-5 Data Editing ..................A-7 Symbologies ..................A-8 Field of View ..................A-16 Working Range ................A-17 Modbus Exception Code ..............A-18 DFS150-485 2D Fixed Scanner User Guide...
  • Page 2: Chapter 1: Overview

    Chapter 1 Overview Delta Fixed Barcode Scanner is capable to read 1D and 2D barcode. The communication interface is RS485 that supports Modbus and Private protocol. The Modbus mode supports parameter setting and barcode transmission, and private mode supports barcode transmission only.
  • Page 3: Chapter 2: Interface

    Chapter 2: Interface Chapter 2 Interface DFS150-485 has a 9-wire cable, beeper and 4 status LEDs as user interfaces. The wire and LED definitions are shown below. Table 1. Cable wire definition Color Function Green Black & White OUT0 Orange...
  • Page 4: Input And Output

    Chapter 2: Interface 2.1 Input and Output DFS150-485 has 1 isolated input and 2 isolated output, both of them share an isolated common ground COM. The input is bi-direction, so that the COM can be connected to either ground or power while output is not connected.
  • Page 5 Chapter 2: Interface Input Circuit Diagram (I/O COM connect to GND)  Figure 3. Input circuit diagram (NPN input) Figure 4. Input circuit diagram (PNP input) DFS150-485 2D Fixed Scanner User Guide...
  • Page 6: Rs485

    Data length: 7 or 8 bits data.  Stop bit length: 1 or 2 bits times.  Table 3. RS485 default setting Item Default 115200 bps Baud Rate 8 bits Data Bits None Parity Stop Bits 1 bit DFS150-485 2D Fixed Scanner User Guide...
  • Page 7 Data Bit Length 0: 8bits 1: 7bits 2: Invalid value 3: Invalid value Baud Rate 0: 115200 bps 1: 57600 bps 2: 38400 bps 3: 19200 bps 4: 9600 bps 5: RESERVED (Invalid value) DFS150-485 2D Fixed Scanner User Guide...
  • Page 8: Led & Beeper

    Figure 5 shows the behavior of LEDs and beeper after power on. Figure 5. Behavior of LEDs after power on Table 4. Beeper Setting Addr. Access Bit Symbol Description Min Max Default Unit 0: Disable 0x0926 [0] BEEP_EN 1(on) 1: Enable DFS150-485 2D Fixed Scanner User Guide...
  • Page 9: Chapter 3: Power On

    If no command is received within T_(startup_comm), ten seconds, it will switch to the scan mode set by user previously. The default scan mode is Slave Modbus ASCII mode. After set all parameters, DFS150-485 provides two ways to update the setting: reboot or send software reset command. The behaviors after T_(startup_comm) are shown below.
  • Page 10: Chapter 4: Operation Mode

    Chapter 4 Operation Mode DFS150-485 provides user with three operating modes and two data types: Master Private, Master Modbus and Slave Modbus modes, and ASCII or RTU. The operation mode and data type can be set by the control register as shown in Scan Mode, Data type and Trigger Source Table 7.
  • Page 11: Master Private Mode

    IN0 to get a falling edge. Figure 8 shows the contents of barcode and receiving data. Figure 8. Master Private Mode barcode sample Figure 9. Master private mode timing diagram DFS150-485 2D Fixed Scanner User Guide...
  • Page 12: Master Modbus Mode (Plc Link)

    Start address: the data were beginning to fill in. Timeout: the slave device should reply within this period. Retry times: if the slave device do not reply within timeout, DFS150-485 will retransmit the same packet no more than retry times. Finished flag: If the finished flag is set to one that indicated a packet of barcode has been transmitted.
  • Page 13 Chapter 4: Operation Mode Figure 11. Master Modbus mode timing diagram Figure 12. Master Modbus mode error situation Table 5. Master Modbus Setting Addr. Access Symbol Description Default Unit 0x900F [0:7] SLAVE_ID PLC_LK_Slave_ID DFS150-485 2D Fixed Scanner User Guide...
  • Page 14: Slave Modbus Mode

    After set the parameters, user can reboot DFS150-485 with software reset command. The details of status of barcode and software reset please refer to Table 6. Registers in Slave Modbus Mode.
  • Page 15 0x8502 [0:6] Reading Length Bytes Bar code length 0x8503 [0:7] Bar code content Barcode Content Words 0x900B [0:7] Slave ID Slave mode ID 0x900A To reboot DFS Software Reset DFS150-485 2D Fixed Scanner User Guide...
  • Page 16: Chapter 5: Trigger

    I/O trigger must refer to OUT0 - ready signal to make a valid trigger. When Input receives falling edge and trigger is bigger than 1ms (refer to Figure 14. Valid trigger), the scan engine will continuous scan until either trigger timeout (30,000ms), or the barcode is decoded successfully. DFS150-485 2D Fixed Scanner User Guide...
  • Page 17: Command Trigger

    Only available in Slave mode. Using Modbus serial command to trigger the scan engine. The scanner continues scanning until a barcode is decoded successfully, or trigger timeout (30,000 ms). Figure 14. Valid trigger Figure 15. Invalid Trigger DFS150-485 2D Fixed Scanner User Guide...
  • Page 18 Description Reserved Trigger Source 0: MANUAL TRIGGER 1: CONTIUNOUS TRIGGER Data Format 0: ASCII FORMAT 1: RTU FORMAT Communication Mode 0: Reserved 1: MASTER PRIVAT MODE 2: MASTER MODBUS MODE 3: SLAVE MODBUS MODE DFS150-485 2D Fixed Scanner User Guide...
  • Page 19 All symbologies enable 0x8001 SYMBOL_EN Symbologies enable 0x8900 IMG_CAP Image capture 0x8500 CMD_TRIG Command trigger 0x8501 BRCD_STS Barcode status 0x8502 BRCD_LEN Barcode length 0x8503 BRCD_CONT Barcode content 0x8800 FW_VER Firmware version 0x900A SW_RST Software reset DFS150-485 2D Fixed Scanner User Guide...
  • Page 20 Access Description Continuous Sensitivity is a numeric range (from 0-20) that increases or decreases the scan engine's reaction time to bar code. 0 is the most sensitive setting, and 20 is the least sensitive. DFS150-485 2D Fixed Scanner User Guide...
  • Page 21: Appendix

    Default Value Access Description The top boundary (from 0-100) of centering window. Table 15. Continuous mode centering windows - Bottom boundary (0x0F41) Bits Default Value Access Description The bottom boundary (from 0-100) of centering window. DFS150-485 2D Fixed Scanner User Guide...
  • Page 22: Continuous Mode Reread Delay

    Poor Quality 1D Barcode is enable, poor quality linear bar code reading is improved, but the scanner’s snappiness is decreased, making it less aggressive when reading good quality bar codes. 0: disable; 1: enable DFS150-485 2D Fixed Scanner User Guide...
  • Page 23: Manual Mode Centering Windows Configuration

    Default Value Access Description The right boundary (from 0-100) of centering window. Table 23. Manual mode centering windows - Top boundary (0x1140) Bits Default Value Access Description The top boundary (from 0-100) of centering window. DFS150-485 2D Fixed Scanner User Guide...
  • Page 24 1: Reverse barcode only, read only inverted barcodes 2: Reverse barcode and standard barcodes Images capture function with the address 0x8900 from the scanner are not reversed. This is a setting for decoding only. DFS150-485 2D Fixed Scanner User Guide...
  • Page 25: Data Editing

    Determine the hex value from ASCII table. Valid range: 0x20-0x7F If you want to remove the prefix of barcode, set the value to (No Suffix) To add blackslash(\), set the value to 5C twice. DFS150-485 2D Fixed Scanner User Guide...
  • Page 26: Symbologies

    Table 34. Code93 enable (0x8005) Bits Default Value Access Description Code93 On/Off 0:Off, 1:On Table 35. Straight 2 of 5 Industrial enable (0x8006) Bits Default Value Access Description Straight 2 of 5 Industrial On/Off 0:Off, 1:On DFS150-485 2D Fixed Scanner User Guide...
  • Page 27 Access Description GS1-128 On/Off 0:Off, 1:On Table 42. Telepen enable (0x800D) Bits Default Value Access Description Telepen On/Off 0:Off, 1:On Table 43. UPC-A enable (0x800E) Bits Default Value Access Description UPC-A On/Off 0:Off, 1:On DFS150-485 2D Fixed Scanner User Guide...
  • Page 28 Table 48. GS1 DataBar Omnidirectional enable (0x8013) Bits Default Value Access Description GSI DataBar-Omnidirectional On/Off 0:Off, 1:On Table 49. GS1 DataBar Limited enable (0x8014) Bits Default Value Access Description GS1 DataBar Limited On/Off 0:Off, 1:On DFS150-485 2D Fixed Scanner User Guide A-10...
  • Page 29 0:Off, 1:On Table 56. MicroPDF417 enable (0x801B) Bits Default Value Access Description MicroPDF417 On/Off 0:Off, 1:On Table 57. GS1 Composite Codes enable (0x801C) Bits Default Value Access Description GS1 Composite Codes On/Off 0:Off, 1:On DFS150-485 2D Fixed Scanner User Guide A-11...
  • Page 30 0:Off, 1:On Table 64. UPC - E0 enable (0x8023) Bits Default Value Access Description UPC-E0 On/Off 0:Off, 1:On Table 65. UPC – E1 enable (0x8024) Bits Default Value Access Description UPC-E1 On/Off 0:Off, 1:On DFS150-485 2D Fixed Scanner User Guide A-12...
  • Page 31 Table 71. Firmware Version (0x8800) Bits Default Value Access Description 15;0 Firmware Version Table 72. Software Reset (0x900A) Bits Default Value Access Description After set the parameters, user can reboot DFS by set the value to one. DFS150-485 2D Fixed Scanner User Guide A-13...
  • Page 32 The register will be effect after the next reboot. Name Reset Access Description Stop Bit Length 0: SHORT STOP BIT (1bit) 1: LONG STOP BIT (2bits) Parity Type 0: NONE PARITY 1: ODD PARITY DFS150-485 2D Fixed Scanner User Guide A-14...
  • Page 33 5: RESERVED (Invalid value) 6: RESERVED (Invalid value) 7: RESERVED (Invalid value) Table 77. PLC Link Slave (Master Modbus Mode) ID (0x900F) Bits Default Value Access Description The Slave ID in master Modbus mode. DFS150-485 2D Fixed Scanner User Guide A-15...
  • Page 34: Field Of View

    Appendix Field Of View Focus Standard Range(SR) 21.2 ± 3.5∘ Horizontal Field Angle (degrees) 16.5 ± 3.5∘ Vertical Field Angle (degrees) DFS150-485 2D Fixed Scanner User Guide A-16...
  • Page 35: Working Range

    Appendix Working Range Barcode Working Range 5mil code39 50~140 mm 20mil code39 45~480 mm 13mil EAN/UPC 50~400 mm 6.67mil PDF417 40~180 mm 10 mil Data Matrix 45~140 mm 20mil QR code 45~245 mm DFS150-485 2D Fixed Scanner User Guide A-17...
  • Page 36: Modbus Exception Code

    If the trigger command is received in the startup communication period or system is in the busy period. INVALID TRIGGER Refer Figure 6. Start-up communication to Scan mode and Figure 14. Valid trigger to create a valid trigger. DFS150-485 2D Fixed Scanner User Guide A-18...

Table of Contents