Max./Min. Menu; Limit Menu; Fig. 7: Max./Min. Menu - Magnet-physik FH 52 Operating Instructions Manual

Gauss-/teslameter
Table of Contents

Advertisement

Gauss-/Teslameter

3.3.10 Max./Min. Menu

The Max./Min. function allows capturing maximum and minimum values.
If Absolute is selected, the maximum is shown in the line below the measured value. The
maximum is the highest reading that was captured since the
this maximum is evaluated, the sign of the reading is not taken into account.
If Max./Min. is selected, the maximum, the minimum and the difference between maximum and
minimum are shown in the line below the actual reading. The maximum is the highest reading
that was captured since the
was captured since the
taken into account.
Reset
The
key deletes the captured values. These are also reset if the device is switched off, if
the signal mode is switched between AC and DC and if the device enters standby. The
Max./Min. function can be used together with the relative function.
The Max./Min. function is designed to observe slowly changing signals. A change in the
magnetic field, that is not visible in the display, cannot be captured. The Max./Min. function
should not be confused with a fast Peak function of more sophisticated measuring instruments.
In the following cases the function is particularly useful:
• Measuring very inhomogeneous fields, for example on the surface of magnets. Here
often only the measurement of the maximum value is sufficiently reproducible.
• To capture readings in hard to get at places.
• To display the maximum field when field orientation is unknown. The probe is rotated
slowly and the largest value will be captured.

3.3.11 Limit Menu

If the Limit function is enabled and the measured value is larger than the upper limit level, High
will be displayed. If the measured value is smaller than the lower limit level, Low is displayed
(see Fig. 3). If the measured value is in between, OK is shown.
BA - Nr.: 1620010102

Fig. 7: Max./Min. menu

Reset
key was pressed last. The minimum is the lowest reading that
Reset
key was pressed last. Here in DC mode the sign of the readings is
Reset
key was pressed last. When
FH 52
Page 20

Advertisement

Table of Contents
loading

Table of Contents