Options (To Be Ordered Separately); Taking Into Operation - Magnet-physik FH 52 Operating Instructions Manual

Gauss-/teslameter
Table of Contents

Advertisement

Gauss-/Teslameter

2.7 Options (to be ordered separately)

The following parts are optionally available for the FH 52:
• Further probes, see also chapter 3.7
• Shielding chamber: a small chamber in which the tip of probe is placed to eliminate
background magnetic fields like the earth's magnetic field during zeroing.
• AC adapter with USB connector (different power line connector styles are available).

2.8 Taking Into Operation

To take the FH 52 into operation, please proceed as follows:
1. Install the batteries: Open the door on the bottom side of the FH 52. Install 4 batteries
observing the polarity for each cell into the battery compartment. Close the door again.
2. Connect the probe: Plug the probe into the socket on the top face of the measuring
instrument. An arrow on the top side of the plug housing indicates the proper plug
position. You can remove the probe again by pulling the plug. Do not try to twist the
plug.
3. Switch on the device: Press the
device name appears. Then the device returns into the operating condition in which it
was when switched off.
The FH 52 should always be switched off before a USB power supply (AC adapter) is
connected or removed. For details see chapter 3.4.
BA - Nr.: 1620010102
On/Off
key. First the Magnet-Physik logo and the
FH 52
Page 13

Advertisement

Table of Contents
loading

Related Products for Magnet-physik FH 52

Table of Contents