8. Use the cursor to select a range of readings that represent 0% oxygen
and click the appropriate Accept reading button;
9. Use the cursor to select a range of readings that represent 100% oxygen
and click the appropriate Accept reading button;
10. To accept and write the calibration coefficients to the logger press the
bottom button.
Hydrotide Druck Pressure Sensor
The calibration procedure for a depth sensor that has resulted in this window is
as follows:
1. Set the logger into a logging mode;
2. Hold the depth sensor out of the water (zero depth) for a sufficient time;
3. Immerse the depth sensor to a known depth for a sufficient time (the 'at
depth' value);
4. Retrieve the data from the logger;
5. Open the View statistics/Two-point calibration window;
6. Select the depth channel using the display cursor;
7. Select the raw data view using the Change data view button;
8. Use the cursor to select a range of readings that represent zero depth and
click the upper Accept reading button;
9. Set the fixed-point value of the bottom set of boxes to the actual depth
value used in step 3.
10. Use the cursor to select a range of readings that represent the 'at depth'
value and click the lower Accept reading button;
11. To accept and write the calibration coefficients to the logger press the
bottom button.
Other calibration details can be found at
http://www.rbr-global.com/technical.htm
The Save fixed-point values for future use button will save the user's
chosen fixed point values to the RBR Windows Software to be used with the
next sensor of the same type.