2) Press "Ready". Now the machine will perform a procedure and just wait until the
machine has completely stopped. Pay attention that the feeding hose is
fully primed! If NOT, then perform step 3-5 below a few times before take
the final reading!!
3) Now press "Test Pump
Accuracy ". A popup window
will now ask you to put the
waste hose into the included
measuring cylinder. Then
press "Test Pump".
4) Water will now be delivered in
the measuring cylinder, keep
the hose in place until the
pump has stopped.
5) Read the water content in the measuring cylinder. A
perfectly calibrated pump will give 7.0 ml.
6) Now its actually possible to redo the test if you are uncertain
you measured or read incorrectly, just press "Test Pump
Accuracy" again and follow instructions again from step 3.
7) When you are done, finish the session by pressing "Empty
Hose". The app will not allow you to leave the page
before you press this of security reason. But if the app,
of any unpredictable reason, not force you to empty
hose, always do that immediately through the
maintenance page!
8) Don't forget to put waste hose back into the container where you collect the
waste water.

Syringe

1) Open the Mastertronic app, and enter Test Pump/Syringe
Accuracy, and choose Syringe
2) Press "Ready".
3) Now press "Test Accuracy" and the syringe will move to a
certain position.
4) Read the syringe very carefully and if perfectly calibrated it
will stop at 0.60 ml.
5) End by pressing "Empty Syringe"

Dashboard

The Dashboard is the Home Screen in the Focustronic app. Here you will get an
overview in graphical form of all the parameters. By choosing one of the headline
14