Syringe; Oli Calibrate; Test Pump/Syringe Accuracy; Stepper Pump (Sample Water Pump) - Focustronic Mastertonic User Manual

Table of Contents

Advertisement

Syringe

1) Open the Mastertronic app, and enter "calibrate", and choose "Syringe"
2) Press "Ready".
3) Now press "Start/Redo" and the syringe will move to a certain position.
4) Read the syringe very carefully (note that every line corresponds to 0.02ml), and
enter the value (for instance 0.64 ml) in the app and save.
5) Now its actually possible to redo the calibration if you are uncertain you read the
syringe perfectly, just press "Start" then "Redo" and take a new reading then
save.
6) Now its calibrated and if it its perfectly calibrated it should go to exactly 0.60 ml
when you check test it with "test syringe".

OLI calibrate

Mastertronic has the possibility to measure the yellow discolouration of the water,
which is a very good indicator of the overall organic load and decomposition levels.
With this parameter you have a very good idea of when to replace any carbon media
or make other actions to reduce the build up of organics.
1) Add around 10 ml of RO or distilled water into the reaction chamber. Be careful
not spill water outside.
2) Open the app, and enter "calibrate", and choose "OLI" and Press "calibrate".
3) The machine will now start to calibrate automatically and after that it will empty
the reaction chamber automatically.

Test pump/syringe accuracy

Tip: Disable screensaver in phone/tablet.
If for some reason you want to check the calibration status/
accuracy of the pumps or syringe, please use this function.

Stepper pump (Sample water pump)

Pump A (RO water pump)

1) Open the Mastertronic app, and enter "Test Pump/
Syringe Accuracy", and choose "Stepper" or "Pump A"
13