System Safety Requirements; Grounding; Power Requirements; Power Supply - NDS 90R0123 User Manual

Radiance ultra 4k/uhd 32"
Table of Contents

Advertisement

System Safety Requirements

External equipment connected to the signal input/output or other connectors of this product for use in a patient
environment must comply with the requirements of ISO and UL/EN/IEC 60601-1 safety standards. A person who
connects such equipment to this product has by definition formed a system, and is responsible for compliance of
that system to the same ISO and UL/EN/IEC 60601-1 safety standards.
To prevent fire or shock hazards, do not expose this unit to rain or moisture.

Grounding

This product is energized from an external electrical power source for 2017/745 Medical Device Regulation of
European Union Class I equipment. It is the responsibility of the installer to test the product's earth ground to
verify that it complies with the hospital, local and national impedance requirements.
A ground post is located on the back of the product to use for grounding the chassis of the unit. Any such
grounding must be installed in accordance with applicable electrical codes.
This equipment/system is intended for use by healthcare professionals only.
This product is a Class I medical device. No modifications are allowed.
This product is intended for continuous operation.

Power Requirements

Power Supply

This product complies to the listed safety standards only when used with the supplied medical grade power
supply:

Power Cord

Use a hospital grade power cord with the correct plug for your power source.
Disconnect the power cord from the AC mains. The power cord is the only recognized disconnect device.
MEDICAL EQUIPMENT should be positioned so that the disconnect device is readily accessible.
Monitors should be powered from a center tapped circuit when used in the US at voltages over 120 volts.

Intended Use and Contraindications

Intended Use
This product is intended for use in a medical environment to display high quality video and graphic images.
Contraindications
1. Do not use this product in the presence of flammable anesthetics mixture with air, oxygen or nitrous oxide.
2. To prevent fire or shock hazards, do not expose this product to rain.
3. No part of this product may come in contact with a patient. Never touch the product and a patient at the
same time.
4. This product is capable of displaying Radiology (PACS) images for reference only, and is not intended for
diagnostic use.
5. For mission critical applications, we strongly recommend that a replacement unit be immediately available.

Image Retention Warning

Warning: Leaving a fixed (constant) image on the monitor for a long period of time can result in image
retention. Avoid leaving a fixed image on the monitor, or turn the monitor off when it is not in use.
2 | Safety Information
Model BridgePower BPM150S24F06
AC Input 100 - 240 Volts, 50 to 60 Hz
DC Output 24 Vdc, 6.25 A, 150 W

Advertisement

Table of Contents
loading

This manual is also suitable for:

90r012590r0132

Table of Contents