Download Print this page
Xylem Ebro TFI 260 Operating Instructions

Xylem Ebro TFI 260 Operating Instructions

Advertisement

Available languages

Available languages

Quick Links

TFI 260 Thermometer Operating Instructions
The thermometer is a non-contact infrared thermometer, and always with Real-Time Maximum (MAX) data.
Please remember to keep away from children and don't use it for safety related applications.
(Default Screen)
1. Simply aim the thermometer at the measure target with Lens (④) and press
Meas. key (⑥) to display the surface temperature.
2. During the measurement, the MAX reading will be displayed beside the
icon.
3. The Distance:Spot is 12:1. Please make sure the target area is within the field
of view.
Functions:
o
o
C or
F
Press Mode key (②) for
Backlight
LCD Backlight: always on.
Laser
ClassIIIa Laser: always enable while measuring.
Memory
Recall the last reading memory by pressing Mode key (②) while power off.
EMC/RFI
Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of approximately
3 volts per meter, but the performance of the instrument will not be permanently affected.
* Multi-Beams
specify the approximate measurement
area for better targeting.
o
o
C or
F transferred.
Xylem Analytics Germany GmbH - ebro
Dr.-Karl-Slevogt-Straße 1 - D-82362 Weilheim
Germany
CAUTION
1. WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAM–PERMANENT EYE DAMAGE MAY RESULT.
2. USE EXTREME CAUTION WHEN OPERATING THE LASER.
3. NEVER POINT THE DEVICE TOWARDS ANYONE'S EYES.
4. KEEP OUT OF REACH OF ALL CHILDREN.
STORAGE & CLEANING
The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times, care should be taken when
cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol. Allow the lens to fully dry before using the
thermometer. Do not submerge any part of the thermometer. The thermometer should be stored at room temperature between –20 to
+65°C (-4~149°F).
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
'Er2' is displayed when the thermometer is exposed to rapid changes in the ambient temperature. 'Er3' is
displayed when the ambient temperature exceeds 0
The thermometer should be allowed plenty of time (minimum 30 minutes) to stabilize to the working/room
temperature.
Error 5~9, for all other error messages it is necessary to reset the thermometer. To reset it, turn the instrument
off, remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If the error
message remains please contact the Service Department for further assistance.
'Hi' or 'Lo' is displayed when the temperature being measured is outside of the measurement range.
BATTERIES
The thermometer incorporates visual low battery indication as follows:
'Battery OK': measurements
'Battery Low': battery needs to be
are possible
replaced, measurements are still possible
When the 'Low Battery' icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries.
Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.
Dispose of used battery promptly and keep away from children.
SPECIFICATION
Measurement Range
Operating Range
Accuracy, (Tobj=15-35°C,Tamb=25°C)
Accuracy (Tamb=23±3°C)
Emissivity
Resolution
Response Time (90%)
Distance:Spot
Battery Life
Dimensions
Weight
** The thermometer will automatically shut off if left idle for more than 15sec.
o
o
C(32°F) or +50
C (122°F).
'Battery Exhausted': measurements are
not possible
o
o
-60 to +550
C (-76 to +1022
F)
0~50°C (32~122°F)
±1.5°C (2.7°F)
Tobj=0~550°C: ±2% of reading or 2°C (4°F) whichever is greater
o
C: ±(2°C+0.05/degree C)
Tobj=-60~0
0.95 fixed
0.1°C/0.1°F at -76 to 999.9(°C/°F), otherwise 1°C/1°F
1sec
12:1
Typ.18, min 14 hours continuous use
114.79 x 162.15 x 40.00 mm (4.52×6.38×1.57 inch)
179.2 grams(6.32 oz) including batteries (AAA*2pcs)

Advertisement

loading

Summary of Contents for Xylem Ebro TFI 260

  • Page 1 3 volts per meter, but the performance of the instrument will not be permanently affected. Response Time (90%) 1sec Distance:Spot 12:1 Xylem Analytics Germany GmbH - ebro Battery Life Typ.18, min 14 hours continuous use Dr.-Karl-Slevogt-Straße 1 - D-82362 Weilheim Germany Dimensions 114.79 x 162.15 x 40.00 mm (4.52×6.38×1.57 inch)
  • Page 2 Stärke von 3 Volt/Meter oder mehr betrieben wird. Die Funktionstüchtigkeit des Geräts wird davon aber nicht dauerhaft beeinträchtigt. Reaktionszeit (90%) 1sec Verhältnis Entfernung:Messfleck 12:1 Xylem Analytics Germany GmbH - ebro Batterielebenszeit Typ.18, min 14 Stunden Dauerbetrieb Dr.-Karl-Slevogt-Straße 1 - D-82362 Weilheim Abmessungen 114.79 x 162.15 x 40.00 mm (4.52×6.38×1.57 inch)