Section 3
Maintenance and Troubleshooting
Following is a list of nontechnical operational checks. These
checks should be performed periodically to ensure proper
equipment operation. If during this procedure, the operator
discovers a malfunction, contact ZOLL Technical Service.
,QVSHFWLRQ
Assure that the unit is clean (with no fluid spills) and nothing
is stored on the unit.
Inspect all batteries for good condition.
&OHDQLQJ
Clean the Base PowerCharger
soap and water.
Do not immerse any part of the Base PowerCharger
water. Do not use alcohol or ketones (MEK, acetone, etc.).
Do not autoclave the Base PowerCharger
2SHUDWLRQDO &KHFNV
1. Connect the unit to live AC mains. All four indicator lights
will illuminate for approximately 4-5 seconds.The
POWER indicator will remain lit while the other lights go
off.
2. With the battery compartment empty, check that all
indicator lights except POWER are off.
3. Press the TEST button to check that the unit is fully
operational.
4. All four indicator lights will illuminate for approximately 4-
5 seconds.
5. If the TESTING light and BATT. READY light remain lit
for three seconds while the other lights go off, the unit is
fully operational without any defects.
6. If the FAULT light remains lit, the unit is defective.
 ,QVHUW D EDWWHU\ LQWR WKH FRPSDUWPHQW 9HULI\ WKDW WKH
&+$5*(5 21 OLJKW LOOXPLQDWHV LPPHGLDWHO\
, with a soft cloth, mild
1x1
in
1x1
.
1x1
7URXEOHVKRRWLQJ
The troubleshooting guide provided below is intended for
use by nontechnical personnel during operation. This
section answers many of the common problems or
questions that arise during operation.
If trouble persists after consulting this guide, call ZOLL and
ask for the Technical Service Department.
1. POWER light does not illuminate.
Check that unit is plugged into an appropriate power
source.
Check for defective power cord.
Plug unit into another power source.
Check for blown fuse.
2. CHARGER ON indicator light does not illuminate when
battery is in compartment.
Check that battery is properly seated in compartment.
Check that battery contacts are clean and not
damaged.
Check that unit is plugged into an appropriate power
source.
Test the battery compartment for functionality.
Place another battery in the compartment.
3. TESTING light does not illuminate when TEST button is
pressed.
Check that battery is properly seated in compartment.
Check that unit is plugged into an appropriate power
source.
Verify that battery compartment is fully operational with
no defects.
Repeat test cycle.
,I EDWWHU\ IDLOV VHFRQG WHVW UHPRYH EDWWHU\ IURP
use.
13