Notes
Example
When repeating the procedure, all references to channel
1
should be
changed to the channel being tested. Turn off all unused channels
using the
BLANk
command.
Bandwidth errors can be caused by the need for adjustment. Perform
high frequency pulse response a&ustment, (see Adjustment
procedures, Chapter
5)
before troubleshooting instrument. If
adjustment fails to correct problem, cause may be the attenuator or A2
PCA See troubleshooting in Chapter 8.
This program is provided to show how to perform a bandwidth test,
and print out the results of each measurement. The program pauses to
allow the user to a&ust the signal generator and power meter.
The example is written using:
an HP-IB select code of 7, primary address of 09, and
secondary address of 05 for the HP E1426A
COMPatible programming language
an HP Series 200/300 Computer with HP BASIC
ecu
t
OUTPUT 70905; "*CLSm
OUTPUT 70905; " ' R S T "
OUTPUT 70905;"TIM:RANG
2E-6"
OUTPUT 70905; "VIEW CHAN1"
OUTPUT 70905;"BLAN CHAN2, CHAN3, CHAN4"
OUTPUT 70905;"CHANl:COUP DCF"
OUTPUT 70905; *'TRIG:SOUR CHANl"
OUTPUT 70905;"ACQ:TYP AVER"
OUTPUT 70905; "ACQ:COUN 32"
OUTPUT 70905;"CHANl:RANG 3.2E-1"
OUTPUT 70905;"MEAS:SOUR CHAN1"
OUTPUT 70905; "DIG CHAN1"
OUTPUT 70905; "MEAS:VPP?'*
ENTER 70905;A
PRINT A
PAUSE
!
To
adjust signal generator and power meter.
OUTPUT 70905;"TIM:RANG
5E-9"
OUTPUT 70905;"MEAS:SOUR
CHAN1"
OUTPUT 70905;"DIG CHAN1"
OUTPUT 70905;"MEAS:VPP?"
ENTER 70905;B
PRINT
B
PAUSE
END
!
To calculate response and change test set up.
!
Repeat
for channels 2-4
.
!
Repeat
for <80 m V range.
February 9, 1991
Verification Tests
4-17