Test failures can be caused by improper cabling, or improper selection
of the interface select code, primary address setting, andlor secondary
address setting. Verify proper connection and address selection using
the applicable command module and mainframe manuals before
troubleshooting.
Example
For the example, use:
an HP-IB select code of 7, primary address of 09, and secondary
address of 05 for the HP E l426A
COMPatible programming language
an HP Series 2W300 Computer with HP
BASIC
OUTPUT 70905; "*CLS8'
OUTPUT 70905; *'*RST"
OUTPUT 70905; "SUMM:PRES1*
OUTPUT 70905; "TEST: TALL"
OUTPUT 70905;"SUMM:QUES:TEST?"
ENTER 70905;A
PRINT A
OUTPUT 70905; "RUN"
END
Ckar status.
Resets the HP E1426A to its default
state.
P ~ s e t s the HP E1426A.
Perfbrm test d l routine
Read t a t event register.
Enter test ewnt register results.
Print t a t ewnt register results.
h t a t t the instrument.
Terminate program.
Note
After a test all is performed, the
RUN
command must be executed
to
restart the instrument.
4-8. Operation
There are no operation verification procedures for the HP E1426A
Verification
Use the Performance verification test procedures for post repair check-
out.
4-9. Performance
The procedures in this section are used
to
test the HP E1426A
Verification
Oscilloscope modules electrical performance using the specifications
in Appendix A of the HP E1426A User's Manual as the performance
standards. These tests are suitable for incoming inspection,
troubleshooting, and preventive maintenance. All tests can be
performed without accessing the interior of the instrument.
Verlflcatlon Tests