Probe-Zero Function
The AD-3255 is equipped with a probe-zero function to calibrate
the zero point of the transducer in the pulse - echo mode.
The sound velocity can be calibrated with a single point calibration,
two point calibration or numerical input.
The coupling state symbol
contacting the material surface.
The AD-3255 can store user settings and calibration data.
The battery indicator
The AD-3255 is equipped with an automatic power-off function:
When no operation is performed for approximately 5 minutes, the
display blinks for 20 seconds and then the power is turned off.
The AD-3255 can measure the thickness of other metal, plastic,
ceramics, composite materials, epoxy resin, glass and a variety of
other materials.
The range of sound velocity:
The range of thickness:
Using the accessory transducer.
indicates the state of the transducer
indicates the battery power.
- 5 -
1000 to 9999 [m/s],
0.8 to 200 [mm],