Overview:
In the off state press the [CAL] button and keep it pressed, press the [ON/OFF] button turn
on the balance. Now the balance is starts in the setting mode. The [CAL] button changes
the setting (C1 to C2). Press the [TARE] button to change the setting value (C1---0 to
C1---1). After all points have been configured, press [CAL] button again. Scale will save
the changes and reboot in weighing mode. the balance automatically into weighing mode.
a) C1 = sensitivity setting - values: 0 1 2 3 4
The lower the value, the faster the reaction speed and sensibility. A higher value
will create a mean measurement of several seconds and improve stability.
Filter 0: Setting for dispensing.
Filter 1/2: Sensitive and fast, for very quiet using location.
Filter 3/4: Robust but slow, for bad environment conditions or as personal scale.
b) C2 = filtering strength setting - values: 0 1 2 3
The higher the value, the higher the stability.
Filter 0: Setting for dispensing, every weight change will be shown in the display.
Filter 1/2: For the most use cases.
Filter 3: Slow but robust, best settings when only bigger items/parcels are placed
on the scale. Tiny disturbances that would change the shown weight in the
display are filtered out and not shown (but also tiny weight changes during
dispensing).
c) C3 = Automatic turnoff - values: 0 1 2 3 4 (only valid for Battery power supply)
0: disabled, scale will stay on. 1: after 15 min. 2: after 30 min. 3: after 45 min.
4: after 60 min.
After C3 press again the "UNIT"-key to save the changes and restart the scale in
weigh-mode.