3.2.4 "COUNT" count key: sampling of article by "COUNT" key to enter the counting state
(1) Choose the quantity of samples: To precisely count the articles, firstly the quantity of
samples for counting should be chosen in according with the weights of the articles. The
quantity of samples can be chosen from "1-10-20-50-100". Those quantity has to be
counted manually. Note here: The smaller the items weight - the bigger the chosen
reference number of items should be to receive a precise result.
(2) Place the chosen numbers of samples on the empty pan, the balance will show the
total weights of samples. Then press "COUNT", the displaying will show "1", at the same
time, the right of the displaying show unit "pcs", with indicate that scale is now in counting
state, and all the samples are one unit. Then press "UNIT" to choose the correct reference
numbers, you can choose between "1, 10, 20, 50, 100". The quantity of articles and
samples must be the same.
(3) Now just add or remove further items, the scale will calculate based on the piece
weight the new amount of items. Press "COUNT" to go back to normal weighing mode.
3.2.5 "UNIT" unit conversion key
Press "UNIT", the units can be changed between "kg". "lb". (1lb=453.59g).
3.2.6 Calibration / Adjustment
NOTE: The scale has been calibrated before shipment and comes ready to use. If the
scale once does no longer work correct and shows wrong results, you can use a new
calibration to adjust the scale again. A calibration can also be done, to adjust the scale as
good as possible to the local environment conditions at the place of use. For a calibration,
a preset adjustment weight is required (see table above). Without the correct adjustment
weight the calibration is not possible and leads to wrong results.
Turn on the scale by pressing [ON/OFF] without a weight on the scale. Wait until the
display shows "0.00".
Press and hold [TARE] key, then short press [UNIT]. The display shows "C XXX" (XXX =
the weight that is needed for this type of scale, like "C 50" for a PSB75 or "C 200 at PSB
300"), Release the buttons
Place the requested weight on the surface and wait until the scale indicates the value.
After about 10 seconds without weight change will show "PASS".
As soon as you see "PASS" the calibration is completed successfully.
4. Setting menu
The settings of the balance can be changed in the setting menu. By the factory the user
menu has been set in a way that normally no more changes must be made, only at special
conditions of use. The balance can be adjusted to individual weighing requirements like a
faster reaction speed during dosages or to filter out more disturbances from bad
environment conditions if you use the scale mobile.