WIRE CONNECTIONS

The housing has single cable entry and can be attached to screw plugs, cord grips or conduit with the ½" NPT
thread. Note: the customer must confirm the suitability of those connectors for the specific application
requirements and cabling; and replace them when necessary. IP66-rated screw plugs and cord grips have to
be properly mounted and tightened around cable of suitable type and diameter to ensure the IP66 rating of the
housing.
Note: Liquid-tight cord grip and ferrite beads are included with the
Ferrite Bead
Liquid-Tight Cord grip
sensor (see Specification Section for cord grip data).
Note: Always include the ferrite bed when using the cord grip or
when using non-metallic conduit.
Note: Always shield the signal wire per instructions on the wiring diagram.
Conduit Connection
Cord Grip (Liquid Tight)
Avoid Condensation in the Conduit
You can give your instrument additional
protection against moisture penetration by
leading the conduit connection or cable
downward in front of the cable entry.
Condensation in the conduit will thus not enter
the sensor enclosure.
QS300840 Rev A2
9
|