Introduction
|
1.2.6 Electromagnetic Compatibility
Important: The ventilator system must be
installed and put into service according to
Electromagnetic Compatibility, SERVO-air
Ventilator System.
In order to ensure that the SERVO-air
Ventilator System, during electromagnetic
disturbances, will deliver ventilation at the
connection port within the alarm limits set by
the user, or generate an alarm condition, the
following essential performance (IEC 60601-1)
has been monitored during electromagnetic
immunity tests:
Delivered volume
Monitoring of:
-
Oxygen concentration
- Airway pressure
- Expired volume
- Internal electrical power source
- Mains power status
- Gas supply
- Gas temperature
Ability to generate alarms
No degradation or failure of the essential
performance has been observed during these
tests.
No other effects on the ventilator system have
been observed during the electromagnetic
immunity tests.