CONTENTS Before the first use............Product Overview.............. Quick Start Guide.............. Output Capacity..............Specifications................ Technical Support.............. Notes...................
BEFORE THE FIRST USE Remove the transport screws under the FLPG Plus. These screws are used to protect the compressor inside during the transport. Please keep these screws in case of future transport.
PRODUCT OVERVIEW The pressure value displayed is in units of bar (1 bar = 14.5 psi) The rear pressure output delivers the same pressure as the front output. One or the other can be used for better tubing management, or both at the same time to pressurize two pressure controllers.
QUICK START GUIDE Please, always place the FLPG Plus on a horizontal and stable surface. Keep clear for at least 10 cm at sides of FLPG Plus. This allows for ventilation and the air drying system to function properly. Start using the FLPG Plus...
Page 6
* The MCFS 2000 mbar range channel is compatible with the 2.3 bar supply pressure, despite its nominal supply pressure of 2.4 bar The FLPG Plus is designed to work at 2.30 bar or below. Do not operate above this value, or the output pressure can become...
Page 7
QUICK START GUIDE Connect the pressure controller's pressure inlet to the FLPG Plus' pressure output (front or/and rear) Power OFF the FLPG Plus after use. Doing so will extend service life of the compressor...
OUTPUT CAPACITY The FLPG Plus' output capacity is the number of channels it can drive simultaneously. This capacity depends on the output pressure set with the pressure setting regulator. When set to 2.1 bar to drive the Flow EZ 2000 mbar channels (or 2.3 bar to drive MFCS 2000 mbar channels), the FLPG Plus can drive up to 4 channels at the same time. When set to 1.1 bar to drive the Flow EZ 1000 mbar channels (or 1.3 bar to drive MFCS 1000 mbar channels), the FLPG Plus can drive up to 8 channels at the same time. When set to lower pressures to drive lower range channels (see tables on page 6 for more information), the FLPG Plus can drive up to 8 channels at the same time. Do not connect more channels than the numbers shown a b o v e .
SPECIFICATIONS FLPG005 Product Number: From 0.1 bar to 2.3 bar Output pressure: 8 Fluigent's 1000 mbar range channels Output capacity*: 8 Fluigent's below 1000 mbar range channels Output capacity*: 4 Fluigent's 2000 mbar range channels 24V DC Voltage: 1.6 A Current: 179 x 154 x 264 mm (W x H x D) Dimensions: 5.9 kg Weight: Indoor clean environment...