3.2.2 Connector
Good, recommended practice is to remove power from any cable prior to connecting or disconnecting it.
1) The CDM900 is provided with one 9-pin D-sub male connector used for the I/O interface. Fabricate a cable to
connect to the vacuum gauge as shown below:
3.2.3 Connector pin-out
9-pin D-sub (DE-9P) male connector (used for analog output and digital I/O)
PIN NUMBER
PIN DESCRIPTION
1
Analog output (Linear 0 to 10 Vdc)
2
Not assigned
3
Not assigned
4
Power Input (13 to 30 Vdc, 0.3 W) protected against power reversal and transient over-voltages
5
Not assigned
6
Not assigned
7
Remote zero adjust
8
Analog output common
9
Power common GND
case
Connector case
InstruTech
Wiring Diagram
Ident
Measurement
signal
10
7
1
8
4
9
case
CDM900 Micro Bee
Remote
Zero
Adjust