IDS-1100 Dermatoscope
The customer or the user of the device should assure that it is used in such environment.
Immunity test
Electrostatic discharge (ESD)
Acc. To IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
Magnetic field IEC 61000-4-8
Comments: UT is the a.c. supply voltage prior to application of the test level.
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that
Immunity test
IEC 60601 test level
Conducted RF
not applicable
Guidance and manufacturer declaration – Electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
IEC 60601 test level
±6kV contact discharge
±8kV air discharge
not applicable
not applicable
not applicable
3 A/m
Guidance and manufacturer's declaration – electromagnetic immunity
it is used in such environment.
Compliance level
not applicable
Compliance level
±6kV contact discharge
±8kV air discharge
not applicable
not applicable
not applicable
3 A/m
Electromagnetic environment Guidelines
not applicable
Electromagnetic environment - Guideline
Floors should be wood, concrete or covered with
ceramic tiles. If floors are covered with synthetic
material, the relative humidity should be at least
30%
not applicable
not applicable
not applicable
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.