7.3.4. Scan the region to be treated:
1.
Tap the region to be treated from the available regions and it will highlight
white showing you the number of lines that have been delivered in the
center of the region.
The energy and treatment line length will be set to appropriate levels for
the entire treatment session unless manually adjusted by the user.
Figure 7.8 DeepSEE Screen
2.
Ensure that the region to be treated has been cleansed thoroughly.
3.
Apply a thin layer of aqueous ultrasound gel to the area to be treated.
W arning: Too much or too little gel will obstruct skin contact could cause
adverse events such as those listed in Section 2.6. Do not use other
lubricants or lotions because they may damage the transducer.
4.
Place the transducer treatment window flush with the patient's skin and
press the See button on the handpiece to begin imaging.
An image of the patient's tissue appears. The green treat line on the
image shows the depth at which treatment will be delivered. Green tick
marks on the ruler show the lateral positions where the coagulative points
will be placed along the horizontal plane. For example, with length set to
25 mm, and a spacing of 1.5 mm (center to center), a treatment line
would have 17 TCPs.
52