Ulthera DeepSEE DS 4 - 4.5 Instructions For Use Manual page 19

Table of Contents

Advertisement

3.3.5. Handpiece
Ulthera DeepSEE Handpiece (UH-2)
3.3.6. I ncluded Accessories
Ulthera System User Access Key (UK-1)
3.3.7. Treatment Controls
Energy, Spacing, Length
See, Treat, Stop, and Cancel
3.3.8. Treatment Parameters
Treat depths: Probe dependent, 1.5 mm to 4.5 mm
Treat Frequency: 4 MHz, 7 MHz, 10 MHz nominal
Treat Energy: Less than 3 J
Treat Power: 75 W maximum at 4 MHz (Energy < 3J)
Treat Time On: 0 to 150 ms, 1 ms resolution, (Energy < 3J)
TCP Spacing: 1 to 5 mm, 1.5 mm standard, 0.1 mm resolution
Treat Line Length: 5 to 25 mm
Treatment output energy accuracy: + 20%
3.3.9. I mage Display
Modes: B-Mode
Ultrasound frequency range: 12-25 MHz
System Dynamic Range: 110 dB total
Instantaneous Displayed Dynamic Range: 53 dB
Scan Lines: 256, at 0.1 mm spacing
Displayed Field of View: 25.6 x 8 mm
3.3.10. Measurement Tools
Distance calipers, 0.1 mm precision, and measurement accuracy ± 5%.
Ulthera, Inc.
65 W maximum at 7 MHz (Energy < 3J)
15 W maximum at 10 MHz (Energy < 1.5J)