Table 6.2 Transducer Energy Lev els
TRANSDUCER
DS 4 - 4.5
DS 7-4.5
DS 7 - 3.0
DS 7 - 3.0N
DS 10-1.5
DS 10-1.5N
Note: Energy level defaults may vary depending on the guidelines
selected prior to performing a treatment. The user has the ability to
adjust these energy settings by using the Energy control described in
Section 7.2.1. If adjusted, the system shall retain the setting across all
regions for that particular transducer and for the duration of the
treatment session.
Note: Removing and re-inserting a transducer within a treatment session
shall put the transducer back to the energy setting used last, not to the
default for the guideline.
Note: Ending the treatment session and starting a new session with the
same or different guideline will put all transducers back to the default
energy settings as prescribed by the guideline.
36
Page
ENERGY LEVELS [J]
LEVEL 4
LEVEL 3
1.20
1.00
1.05
0.90
0.45
0.35
0.45
0.35
0.25
0.20
0.25
0.20
LEVEL 2
LEVEL 1
0.90
0.75
0.75
0.66
0.30
0.25
0.30
0.25
0.18
0.15
0.18
0.15
LEVEL 0
0.00
0.00
0.00
0.00
0.00
0.00
Ulthera, Inc