Cautions
• Do not install the unit in the following locations:
- Locations with dust, salt, or sulfur.
- Locations directly exposed to sunlight for extended periods of time (in particular, do
not leave in direct sunlight or near a source of ultraviolet light for extended periods,
as ultraviolet light will cause deterioration of the LCD).
- Locations subject to vibration or shock.
- Near heaters.
• Do not use in a location with noise-emitting equipment such as a room with MRI,
CT, X-ray or HF surgical equipment or an operating room. Noise from the equipment
may interfere with the operation of the unit.
Before use / during use
Warnings
• The unit complies with the EMC (Electro Magnetic Compatibility) standard
(IEC60601-1-2). As such, it can be used simultaneously with multiple medical
instruments. However, if instruments that generate noise such as an electric scalpel
or a microwave therapy device are near the unit, check the operation of the unit
during and after use of these instruments.
• If an error occurs or a measurement result is questionable, check the vital signs of
the patient by auscultation or palpation. Avoid relying solely on the measurement
results of the unit when judging the patient's condition.
• Only trained healthcare providers should use this device. Do not allow patients to
operate this device.
• Properly connect the connectors and AC adapter cable.
• Do not place objects or liquids on top of this unit.
• Check the following before using the unit:
- Make sure the AC adapter cable is not damaged (wires are not exposed or broken),
and the connections are firm.
• For the AC adapter connected to the unit, supplies, and optional devices, use only
the standard accessories or OMRON-specified products. This may damage and/or
may be hazardous to the device.
• Do not use in a location with moisture, or a location where water may splash on the
unit.
• This unit is intended for use in physicians' office, hospitals, clinics and other medical
facilities.
• Do not use the unit if it emits smoke, an abnormal odor, or abnormal noise.
• Do not bring cellular telephones or transceivers into the room where the unit is
installed or being used.
• Do not connect multiple monitors to the same patient.
• Do not connect the unit to a power outlet that is controlled by a wall switch.
Cautions
• Before using the unit, verify that none of the following apply to the patient:
- Poor peripheral circulation, noticeably low blood pressure, or low body temperature
(there will be low blood flow to the measurement position)
13
PM-1794-01-06/2018