Tektronix 1502C Service Manual page 162

Metallic time-domain reflectometer
Hide thumbs Also See for 1502C:
Table of Contents

Advertisement

Maintenance
Leading edge not at
Check chart recorder
No
CONTINUE
zero, wrong length,
or other extra
3
or
function modules.
with discontinuities.
Yes
±
Check for <
4 volts
TP4020 and TP4021
on main board.
No
Yes
Check digital delay
Check analog delay
Run Calibration
on main board.
on main board.
Performance Check.
7–15
1502C MTDR Service Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com