HP 3400B User's And Service Manual page 26

Rms voltmeter
Table of Contents

Advertisement

Chapter 2 Calibration Procedures
Performa nce Verification Proced ures
Scale
1 HP
Tenth
Test
2
3
4
1\
3400 B :
Select the
Calibrator: Select a
External Multimeter: Select the 10 Vdc range with
resolution.
Verify that the de voltage measured at the rear-panel de output
jack is within the limits shown in the table below.
Continue checking the de output as you adjust the calibrator's
output voltage and frequency to each of the remaining values
shown in the table below.
HP 3400B
calibrator
Range
Output Vo�
I V
0.1 V
I V
0.1 V
I V
0.1 V
0.1 V
I V
0.1 V
I V
I V
0.1 V
I V
0.1 V
24
1
volt range.
O. 1
volt ac signal at
10 Hz.
calibrator
Output Frequency
1 0 Hz
so
Hz
1 MHz
2 MHz
3 MHz
10 MHz
20 MHz
51-\!
digits of
DC Output
(DMM Reading)
-<l . 1 5 V to -<l . 05 V
-1.075 V to -<l .0925 V
-1.075 V to -<l .0925 V
-<l .1 2 V to -<l .08 V
-<l .1 3 V to -<l .07 V
-<l .1 5 V to -<l .05 V
-<l .1 5 V to -<l .05 V
-,