Operation Verification; Performance Verification - Agilent Technologies 75000 Series Service Manual

Summing amplifier/dac module
Hide thumbs Also See for 75000 Series:
Table of Contents

Advertisement

Operation
Verification
Performance
Verification
2-4 Verification Tests
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Operation Verification is a subset of the Performance Verification tests
that follow. For the amplifier, Operation Verification consists of the
following tests:
Test 2-3: Low-level Outputs Test
Test 2-4: Main Output Test
Test 2-5: Offset DAC Test
The procedures in this section are used to test the amplifier's electrical
performance using the specifications in Appendix A of the Agilent
E1446A User's Manual as the performance standards. These tests are
suitable for incoming inspection, troubleshooting, and preventive
maintenance. The results of the Performance Verification tests should
be recorded in the Performance Test Record (Table 2-4).
Performance Verification includes the following tests:
Test #
2-1
Input Attenuation Test
2-2
Output Attenuation Test
2-3
Low-Level Outputs Test
2-4
Main Output Test
2-5
Offset DAC Test
2-6
Bandwidth Test
Test Name