HyQuest Solutions CMC20A Operating Manual

HyQuest Solutions CMC20A Operating Manual

Current meter counter

Advertisement

Quick Links

Hyquest Solutions Pty Ltd 
OPERATING MANUAL
CMC20A
Current Meter Counter
QUALITY SYSTEM
ISO: 9001
CERTIFIED
HYQUEST SOLUTIONS PTY LTD
48-50 Scrivener St, Warwick Farm, NSW 2170, AUSTRALIA
Phone: +61 2 9601 2022 Fax: +61 2 9602 6971
Email:
sales@hyquestsolutions.com.au
Web:
www.hyquestsolutions.com.au
CMC20A instruction manual                                       1 
 Issue 2:19/5/2016 
©
Copyright 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CMC20A and is the answer not in the manual?

Questions and answers

Summary of Contents for HyQuest Solutions CMC20A

  • Page 1 Hyquest Solutions Pty Ltd  OPERATING MANUAL CMC20A Current Meter Counter QUALITY SYSTEM ISO: 9001 CERTIFIED HYQUEST SOLUTIONS PTY LTD 48-50 Scrivener St, Warwick Farm, NSW 2170, AUSTRALIA Phone: +61 2 9601 2022 Fax: +61 2 9602 6971 Email: sales@hyquestsolutions.com.au Web: www.hyquestsolutions.com.au CMC20A instruction manual                                       1   Issue 2:19/5/2016 ...
  • Page 2 CONTROLS ..............................5 ................................5 TART ................................5 ............................5 OTARY WITCH ..............................6 ISPLAY ......................6 ETTING ONTACT EBOUNCE CMC20A F ..................6 ISPLAYING THE IRMWARE EVISION ANNUAL SERVICING ..........................6 FAULT FINDING ............................7 CMC20A instruction manual                                       2   Issue 2:19/5/2016  © Copyright ...
  • Page 3 - 6 AA Alkaline batteries. The actual battery voltage is displayed for 2 seconds each time the unit is switched on. Low voltage is also indicated by all decimal points showing on the display at 5.8 volts. CMC20A instruction manual                                       3   Issue 2:19/5/2016  © Copyright ...
  • Page 4 An internal buzzer sounds at each contact closure during the timing period to indicate correct operation. The display on the CMC20A gives continuous indication of the pulse count during the course of the reading period - and indicates with a resolution of 1 to a maximum of 9999 counts.
  • Page 5 Stop Stops further counting - and allows the counter to be reset and restarted by the Start switch. Rotary Switch Selects a range of preset time periods and the integrate mode for use with ground-feeler weights. CMC20A instruction manual                                       5   Issue 2:19/5/2016  © Copyright ...
  • Page 6 When satisfied with the setting, rotate the Rotary switch one click in any direction. This will save the selected debounce time for all future use. Note that this is saved even when the unit is switched off, or batteries are removed. Displaying the CMC20A Firmware Revision To display the Firmware Revision : ...
  • Page 7 5.8volts. (When unit is switched on, the battery voltage is displayed for 2 secs.) As meter gets faster, the count The contact debounce time is Reduce the contact debounce stops. set too long. time.   CMC20A instruction manual                                       7   Issue 2:19/5/2016  © Copyright ...