EMC GUIDANCE
Voltage dips, short
Not applicable
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
30 A/m
50Hz/60Hz
Power frequency
magnetic field
IEC 61000-4-8
Conduced RF
Not applicable
IEC61000-4-6
10 V/m
Radiated RF
80 MHz – 2,7 GHz
IEC61000-4-3
80 % AM at 1 kHz
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
Table 3
Guidance and manufacturer's declaration - electromagnetic Immunity
Radiated RF
Test
Band
Service
(MHz)
IEC61000-4-3
Frequency
(MHz)
(Test
specifications
for
385
380-390
TETRA
ENCLOSURE
400
PORT
IMMUNITY to
GMRS 460,
RF wireless
FRS 460
communicatio
450
430-470
ns equipment)
710
704-787
LTE Band
13,17
745
780
24
Not applicable
30 A/m
50Hz/60Hz
Not applicable
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Distance (m)
IMMUNITY
Modulation
Modulation
(W)
TEST
LEVEL
(V/m)
1.8
0.3
Pulse
modulation b)
18Hz
FM c) ± 5kHz
2
0.3
deviation 1kHz
sine
Pulse
modulation b)
0.2
0.3
217Hz
810
870
930
1720
1845
1970
2450
5240
5500
5785
27
28
9
GSM
Pulse
2
800-960
800/900,
modulation b)
TETRA 800,
18Hz
iDEN 820,
CDMA 850,
LTE Band 5
GSM 1800;
Pulse
1700-
2
modulation b)
CDMA 1900;
1990
GSM 1900;
217Hz
DECT;
LTE Band 1,
3,
4,25; UMTS
Pulse
Bluetooth,
2
2400-
modulation
WLAN,
2570
217 Hz
802.11
b/g/n, RFID
2450, LTE
Band 7
5100-
WLAN
Pulse
0.2
5800
802.11
modulation
a/n
217 Hz
EMC GUIDANCE
0.3
28
0.3
28
0.3
28
0.3
9