Some materials are more susceptible to corrosion than other. Also, some fluids and additives are more corrosive than others. OFITE offers a variety of cell materials for different levels of corrosion resistance and cost. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 7
Oil on the floor is very slippery and can cause falls and injury. Oil spills on the bench can accumulate and become a fire hazard. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
21. Screw the cell cap assembly into the cell body and tighten it completely. 22. Connect the LCM bottom cap assembly to the cell cap assembly. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 9
42. Disconnect the inlet pressure manifold from the inlet pressure assembly. 43. Remove the cell from the heating jacket and set it upright to disassemble. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 10
50. Disconnect the pressure manifold from the inlet pressure assembly. 51. Remove the inlet pressure assembly and inlet cell cap assembly. 52. Clean and dry the entire cell assembly. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
7. Push in the red knob on the heating jacket. This moves the cell rest plunger into position to support the cell. 8. Place the cell into the heating jacket with the inlet side pointing down. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 13
Make sure the quick-connect fittings are completely engaged. Pull down hard on the ring on the female fitting until it clicks into place. Quick-Connect Fitting Inlet Ball Valve Inlet Pressure Manifold OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 14
23. Place a ceramic disk or a circle of filter paper on top of the o-ring. 24. Screw the cell cap assembly (cell cap and male VCO fitting) into the cell body and tighten completely. Cell Cap Assembly OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 15
33. Hold the outlet pressure assembly with one hand and pull the stop on the heating jacket out of the way to lower the cell fully inside the heating jacket. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 16
35. Place a metal dial thermometer (#154-20) in the top of the cell in the small hole. T-Screw Needle Valve (#170-32) Safety Pin (#171-23-1) Pressure Receiver OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Maintain the pressure on the fluid until the tempera- ture has stabilized. The heating time of the sample should never exceed one hour. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 18
10. Reconnect the inlet pressure manifold to the inlet pressure assembly (quick connect fitting) 11. Unscrew and remove the outlet cell cap assembly. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 19
Also make sure to clean the outlet valve assembly. Inspect all o-rings and replace any that show signs of damage or wear. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
“zero mark” at the very edge of the ruler is useful here. Cake descrip- tions may be subjective and such notations such as hard, soft, rubbery, and fine, etc. convey adequate information on cake quality. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Make sure it engages completely around the circle. The cap should turn freely inside the locking ring. Retaining Ring Locking Ring O-ring OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
5. Place two o-rings (#171-900-025) in the o-ring grooves around the out- side of the piston. Rupture Disk Set Screw O-ring O-ring OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Replace the thrust plate with the curved edges downward. Place a new rubber diaphragm inside the spring case. m. Replace the spring case and hand tighten. Diaphragm Repair Kit (#143-00-1) (#143-00-1) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided. OFITE makes no other warranties or guarantees to the buyer, either express or implied, and the warranties...