300°F (149°C). Anyone running tests above 350°F (177°C) must substitute a complete set of o-rings after each and every test. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 7
Everyone’s consumable requirements will be different, and replace- ment quantities needed will depend upon the number of test performed on a daily and/or weekly basis. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
HTHP Filter Press and especially when one is going to extreme temperatures and pressures. Above 350°F (176°C) all o-rings must be replaced after each and every test. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 12
HTHP Filtration. Under high temperatures and pressures, Nitrous Oxide can detonate in the presence of grease, oil or carbonaceous material. Nitrogen Bottle (#170-37) High Pressure Nitrogen Cylinder Optional (#171-31) Optional OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 13
Viton o-rings is 400°F (204°C). OFITE suggests that all o-rings (cell caps, cell bodies, and valve stem) are changed each time after running a test of 350° F (176° C) or above. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 14
Some materials are more susceptible to corrosion than other. Also, some fluids and additives are more corrosive than others. OFITE offers a variety of cell materials for different levels of corrosion resistance and cost. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 15
9. Do not purge oxidizing or flammable gases in the presence of flame, lit cigarettes, or other sources of ignition or towards people. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 16
All OFITE HTHP cell assemblies are made of Stainless Steel. For very corrosive environments cell assemblies made of Hastelloy C-276 or Inconel are available and recommended. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
20. Connect the back pressure receiver to the outlet valve stem. Lock it in place with the retaining pin and lanyard. 21. Check the needle valve on the outlet end of the receiver is closed (tightened OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 20
39. Release the bottom pressure from the regulator by opening the receiver needle valve and collect any remaining filtrate. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 21
53. Report the description of the filter cake: thin, hard, rubbery, etc. 54. Thoroughly clean all equipment and replace o-rings if the test was conducted above 300°F (150°C). OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
22. Be careful not to spill fluid on the o-ring inside the cell. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 23
9. Place a circle of filter paper on top of the cell o-ring and gently push the paper downward so it contacts the o-ring without the paper binding or pinching. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 24
Thermocouple Assembly (#171-45-1) is available which will directly and accurately measure the fluid temperature inside the cell. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 26
26. Turn the regulator t-screws on both the inlet and back pressure receiver counterclockwise until the t-screw feels “loose”. This will close off the flow of pressurized gas. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 27
There should be a noticeable release of pressure from the cell. Stand for HTHP Cell Assembly (#171-190-028) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 28
After each test, the HTHP filter press should be left so that a new test may be performed with no clean-up and a minimum of assembly. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
“zero mark” at the very edge of the ruler is useful here. Cake descriptions may be subjective and such notations such as hard, soft, rubbery, and fine, etc. convey adequate information on cake quality. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Back Pressure Receiver – Similar to “B”. Hold assembly by the gauge and do not allow it to get wet. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 31
Concoa Regulator (805-1179) both take the same Regulator Repair Kit. Always replace (order separately) the rubber diaphragm, #143-00-1, which is not supplied with the #143-07 kit. Diaphragm Repair Kit (#143-00-1) (#143-00-1) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 32
11. Replace the thrust plate with the curved edges downward. 12. Place a new rubber diaphragm inside the Soring case. 13. Replace the spring case and hand tighten. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided. OFITE makes no other warranties or guarantees to the buyer, either express or implied, and the warranties...