Note:
You can turn the sound off by selecting Sound Off in the set up menu.
This does not affect the 'Lead Off' alarm.
3.14 Error and Warning Displays
1.
You will see this error message if you are trying to use Extended
Recording Mode with a six electrode cable. Either
select Start. . . for Standard Recording Mode or
change the patient cable.
2.
You are trying to use Standard Recording Mode
with a two electrode cable. Either select Start Week
for Extended Recording Mode or change the patient
cable.
3.
There may be a fault in the cable, or you may
require a firmware upgrade to use this cable type.
4.
If ***Error 0007*** message is displayed, contact the
Service Department at Spacelabs Healthcare.
5.
This message is displayed if a patient loses a lead during an Extended
Recording. The Lifecard will also generate an alarm
which will consist of a series of 'ding-dong' tones
lasting 15 seconds. Once the patient has replaced
the lead, the clock display resumes.
6.
This warning is generated if you have selected Start
Week... and the battery is partially discharged.
Note:
7 day recording needs a new alkaline battery.
7.
You will see a warning of this format if you have
selected Start Week.... and the capacity is too
small to record 7 days. Select Continue only if
the recording description meets or exceeds your
requirements
8.
This message appears if the card capacity and
patient cable are not appropriate to provide a 24
hour recording.
9.
The Lifecard has been closed without a Flashcard
inserted. Reopen the recorder and insert the card.
10. The Flashcard has been inserted before being correctly initialized.
Remove the card and follow the instructions to condition and initialize CF
cards.