Appendix 3 Troubleshooting; A3.1 No Display; A3.2 Noise; A3.3 Recorder Error - EDAN CADENCE II User Manual

Fetal monitor
Table of Contents

Advertisement

Appendix 3 Troubleshooting

A3.1 No Display

Symptom
Power indicator is off

A3.2 Noise

Symptom
Noise

A3.3 Recorder Error

Symptom
Paper jam
Recorder does not work
CADENCE II Fetal Monitor User Manual
Possible Cause
Power cable is loose and no
battery
The fuse(s) is (are) blown.
The battery is low.
Possible Cause
The FHR volume is set to be
too high.
Interfered
by
handset
other interfering source
Possible Cause
Wrong
feeding
paper
paper is affected by damp.
The PRINT key is disabled
Out of paper
Ultrasound
transducer
DECG cable connector is
unplugged.
- 62 -
Solution
Tighten the power cable and
check the fuse, or check the
battery
Replace the fuse(s).
Connect the monitor to AC power.
Solution
Adjust the volume down
or
Keep
the
handset
interfering source far away
Solution
or
Feed paper correctly and keep
paper from moist
Press the PRINT key again
Feed paper
or
Connect
with
transducer
or
connector
or
other
ultrasound
DECG
cable