4.2 MECHANICAL OVERRIDE KEY LOCK

In the event of a power or system failure there is a mechanical override key lock that can be used to gain access to the contents of
drawers. The mechanical override is a two-stage process.
1.
First, insert the override key into the override lock on the front of the housing. Turn the key counter-clockwise.
2.
This will eject the cassette handles from the housing. Pulling the handles outwards will allow the cassettes to be removed from
the housing.
3.
Second, use the same override key in the lock on the rear panel of the cassette. Turn the key counter-clockwise to unlock all the
drawers in the cassette. Drawer may now be removed from the cassette.
NOTE: Once either override lock has been actuated, you must turn the key back to the vertical position to reset the mechanical
override, otherwise the system will remain in override mode and normal operation through the MedLink software will not be possible.