Agilent Technologies 100 Operator's Manual page 32

Automated disintegration apparatus
Hide thumbs Also See for 100:
Table of Contents

Advertisement

4
Operating the 100
NOT E
32
The example on page 31 indicates the previously programmed test
length for Position 1 was 30 minutes.
• To accept the current test length, press ENTER.
• To change it, enter the new value in hhmmss format and press
ENTER.
• To disable Position 1, enter 000000 for the test length and press
ENTER.
2 In the display, the numeral 2 flashes, alternating with the previously
programmed delay time for Position 2 in hhmmss format. The delay
time is the amount of time between the start of the program and the
start of dipping for Position 2. For example, you may see the
following alternating displays:
POSITION 3
2
The above example indicates the previously programmed delay time for
Position 2 was 5 minutes.
• To accept the current delay time, press ENTER.
• To change it, enter the new value in hhmmss and press ENTER.
• If no delay is desired, enter 000000 and press ENTER.
There is no delay time for Position 1.
3 In the display, the numeral 2 flashes again, this time alternating with
the previously programmed test length for Position 2. For example,
you may see the following alternating displays:
POSITION 3
2
POSITION 3
000500
POSITION 3
004000
100 Operator's Manual