Mounting Baskets and Filling Beakers

Draining the Water Bath

WA RNING
100 Operator's Manual
1 Ensure the water bath cover is in place on the water bath.
2 Place one, two, or three 1000 mL USP disintegration beakers through
the water bath cover holes so they are suspended in the water bath.
3 Hold the beaker(s) down and pour 900 mL of the required
disintegration medium in each beaker. When full, the beakers sit
securely in the cover and do not float.
4 Place one, two, or three basket racks onto the horizontal basket
supports.
5 Verify the water level in the water bath is above the media level of all
beakers. If not, either add additional water or drain excess water until
the proper working level is reached.
Disconnect the 100 from the AC power receptacle before
draining the tank.
1 Disengage the quick-connect fitting on the left side of the tank and
attach the supplied drain hose.
2 Open the drain valve and drain the water bath into an appropriate
receptacle or sink.
3 Use a clean paper towel or damp cloth to wipe out the water bath and
remove any remaining water from the bottom of the tank.