No. 0 Phillips screwdriver
Cleaning the jumper connector:
1. Apply alcohol to a small area of a lint-free towel and rub the end of the ferrule over the wet
area.
2. Wipe the ferrule on a dry area of the lint-free towel.
3. Using the dusting gas or compressed air, blow the end of the ferrule.
4. Apply the alcohol or ethanol to a lint-free pipe cleaner or swab and wipe off the remaining
parts of the connector.
5. With the other end of the pipe cleaner or swab, dry the areas cleaned.
6. Using the dusting gas or compressed air, blow the areas cleaned.
Cleaning the mating sleeve:
1. Blow compressed air inside the opening of the mating sleeve to remove trapped particles
(Figure 3).
2. Moisten a fresh pipe cleaner with alcohol. Gently clean the opening by inserting the pipe
cleaner using a rotating motion.
3. Blow compressed air inside the opening to remove any remaining material.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figure 2: Typical jumper connector