EARTH/GROUND AT ALL TIMES TO AVOID RISK OF ELECTRIC SHOCK OR
ELECTROCUTION.
The power supply cable and connector must be correctly rated for the power
supply. Only use power cables supplied by GE specifically for the ADTS405.
AC Single
phase
(standard)
DC
(option
405F)
2.1. AC Power supply connection
European Colour
Green/Yellow
2.2. AC Fuses
Two 20mm High Breaking Capacity AC fuses are connected in the live and
neutral supply circuit and protect the entire unit. The AC fuses are rated at
T5AH 250V. On the ADTS 405F fuses are located in the holders mounted on the
pump rack front panel (see Figure 5). On the ADTS 405R fuses are located in
the holder mounted on the power input connector on the rear of the rack (see
Figure 3).
2.3. DC Power supply connection (option)
European Colour
2.4. DC Fuses (option)
On the ADTS 405F, two 1.25" Time Lag fuses are located in the holders
mounted on the pump rack front panel and protect the entire unit.
On the ADTS 405R, DC overcurrent protection fusing is internal and not user
serviceable.

[EN] English

WARNING
THE ADTS MUST BE CONNECTED TO A SUPPLY PROTECTIVE
100/120/230 Vac, 50/60Hz
115 Vac, 400Hz
100/120/230 Vac, 50/60Hz
115 Vac, 400Hz
16 to 30 Vdc
Brown
Blue
Grey
White
Wire Braid
500 VA Max – ADTS 405F
400 VA Max – ADTS 405R
500 VA Max – ADTS 405F
US Colour
Black
White
Green
US Colour
Gray
White
Wire Braid
3
Function
Live
Neutral
Protective Earth (Ground)
Function
+28V
0V Return
Protective Ground
124M8686 Revision A