Recommended Separation Distances between Portable and Mobile RF Communications Equipment and This Device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of
the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum Output
Power of Transmitter
(Watts)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.
46
Separation Distance According to Frequency of Transmitter (meters)
150 kHz to 80 MHz
d = 1.2
0.12
0.38
1.20
3.80
12.00
80 MHz to 800 MHz
d = 0.35
0.04
0.11
0.35
1.11
3.50
800 MHz to 2.5 GHz
d = 0.7
0.07
0.22
0.70
2.21
7.00