Circuit Disconnect Alarm
This setting enables or disables the circuit disconnect alarm. If enabled, an audible alarm will sound when a
large, continuous air leak (such as mask removal) has been detected in the circuit. Select Off to disable the
alarm, or choose 15 or 60 seconds. Selecting 15 or 60 means that the alarm will sound after the circuit has been
disconnected for that amount of time.
Apnea Alarm
This setting enables or disables the apnea alarm. If enabled, an audible alarm will sound when an apnea is
detected. Select Off to disable the alarm, or increase or decrease the setting from 10 to 30 seconds in 10 second
increments. For example, a setting of 10 means that the alarm will sound if the time between spontaneous
breaths exceeds 10 seconds.
Low Tidal Volume Alarm
Select On to enable or Off to disable the Low Tidal Volume alarm. When the alarm is enabled, an audible
indicator sounds if target tidal volume can't be reached.
Low Minute Ventilation Alarm
This setting enables or disables the Low Minute Ventilation alarm. The alarm activates when the calculated
minute ventilation is less than or equal to this setting. Select Off to disable this alarm, or increase or decrease
the setting from 1 l/min to 99 l/min in increments of 1.
High Respiratory Rate Alarm
This setting enables or disables the High Respiratory Rate alarm. The alarm activates when the measured
respiratory rate reaches or exceeds this setting. Select Off to disable this alarm, or increase or decrease the
setting from 1 BPM to 60 BPM in increments of 1.
Changing Options
1. From the Main Menu screen, use the Up/Down key to highlight the Options item.
2. Press the Right key to select Options.
The following settings are available on the Options menu.
Detailed View
Turn Detailed View On or Off using this setting. Detailed view displays additional therapy information on the
Monitor screen.
Language
Select the Language that the software will appear in (English, French, German, etc.). The information on the
screens will display in the language selected here.
Pressure Units
Select the pressure units that will display on the screens. You can choose either cm H
units on the screens will display in the unit of measure selected here.
Breath Indicator
Select Patient or Machine to choose whether the breath indicator flashes on-screen during a patient-triggered
breath or a machine-triggered breath. The default is Machine.
Keypad Lock
Select On to enable or Off to disable the Keypad Lock feature.