PHILIPS Respironics OmniLab Advanced + Clinical Manual page 27

Table of Contents

Advertisement

Setting
Flex Type
Flex
CPAP
Maximum Pressure
Minimum Pressure
Ramp Time
Ramp Start Pressure
The following list gives more detail on the Settings and Alarms listed in the previous tables.
Mode
Change the Mode setting to one of the following therapy modes:
• CPAP
• T
• S
• PC
• S/T
• Auto-CPAP
Flex Type
This allows you to choose the Flex comfort feature setting.
Flex
Set Flex to 1, 2, or 3 to enable the setting. A setting of 1 provides a small amount of pressure relief, with higher
numbers providing additional relief. Select Off to disable the setting.
AVAPS
Select On or Off to enable or disable AVAPS.
CPAP
Increase or decrease the CPAP pressure setting from 4 to 20 cm H
IPAP
Increase or decrease the Inspiratory Positive Airway Pressure (IPAP) from 4 to 30 cm H
0.5. You cannot set the IPAP setting lower than the EPAP setting. IPAP is limited to 25 cm H
feature is enabled.
IPAP Max Pressure
Increase or decrease the setting from 4 to 30 cm H
equal to or greater than the IPAP Min value.
IPAP Min Pressure
Increase or decrease the setting from 4 to 30 cm H
equal to or greater than the EPAP value, and it must be less than or equal to the IPAP Max Pressure.
CPAP
C-Flex, C-Flex+
Off, 1, 2, 3
4-20 cm H
O
2
(0.5 cm H
O res)
2
n/a
n/a
0-45 min
(5 min res)
4-(CPAP-0.5) cm H
O
2
(0.5 cm H
O res)
2
• Auto-BiLevel
• autoSV
• AVAPS-AE
O in increments of 0.5. The IPAP Max Pressure must be
2
O in increments of 0.5. The IPAP Min Pressure must be
2
Mode
Auto-CPAP
C-Flex, A-Flex
Off, 1, 2, 3
n/a
Min Pressure-20 cm H
(0.5 cm H
O res)
2
4-Max Pressure cm H
(0.5 cm H
O res)
2
0-45 min
(5 min res)
4-Min Pressure cm H
(0.5 cm H
O res)
2
O in increments of 0.5.
2
O in increments of
2
O
2
O
2
O
2
O when the Flex
2
23