PHILIPS Respironics OmniLab Advanced + Clinical Manual page 39

Table of Contents

Advertisement

Alarm
Priority
Low Battery
Escalates
(When Battery
from
is Attached)
Medium to
High
AC Power
Medium
Disconnected
(When Battery
is Attached)
Stuck Key
Low
Insert SD
Low
Card (When
Oximeter is
Attached)
Informational Messages
Message
Priority
Card Error
Info
Heated Tube
Info
Failure
Humidifier
Info
Failure
Audible
Visual Indicators
◊◊◊
Medium Priority -
(Medium -
Yellow flashing button;
when approx.
Low External Battery
20 minutes
message appears in
remains)
yellow; on status panel,
box around battery is
yellow.
◊◊◊ ◊◊
High Priority - Red
(High - when
flashing button; Low
approx. 10
External Battery
minutes
message appears in
remains)
red; on status panel,
box around battery
is red.
◊◊◊
Yellow flashing
button; AC Power
Disconnected
message, and a box
appears around
battery in use.
◊◊
Solid yellow button;
Keypad Stuck message.
◊◊
Solid yellow button;
Insert SD Card
message
Audible
Visual Indicators
Card Error message
None
Flashing icon:
None
Flashing icon:
Device Action
Operates
Switch to an alternate battery, or
switch to AC power.
Switches to
Check AC power adapter and
alternate
reconnect it if it has become
power source
disconnected. Make sure device is
not connected to an overloaded
AC circuit.
Operates
Check the keys to determine if
they are lodged in the case. If alarm
continues, place patient on alternate
source of therapy and contact
product support.
Operates
Insert an SD card into the device or
remove the oximeter.
Device Action
Operates
Remove SD Card and use another
card, if available. Ensure card meets
specifications. If condition persists,
contact product support.
Device
Tubing may be overheating or
operate;
malfunctioning. Turn off airflow
Humidifier
and reconnect the heated tubing
shuts down
to the humidifier according to the
humidifier instructions. If the alert
continues to occur, contact product
support.
Device
Humidifier failure. Turn off airflow
operate;
and reconnect the humidifier to the
Humidifier
device according to the humidifier
shuts down
instructions. If the alert continues to
occur, contact product support.
User Action
User Action
35