Process Sensors Sensor Therm METIS M311 Operating Manual

Process Sensors Sensor Therm METIS M311 Operating Manual

Pyrometers
Hide thumbs Also See for Sensor Therm METIS M311:

Advertisement

Quick Links

Pyrometers
METIS M311 / M322
METIS H311 / H322
With 12-pin connector, LED or OLED display and push button device configuration
Versions with integrated optics, optical fiber, PID controller, Profibus / Profinet
Operating Manual

Advertisement

loading
Need help?

Need help?

Do you have a question about the Sensor Therm METIS M311 and is the answer not in the manual?

Questions and answers

Summary of Contents for Process Sensors Sensor Therm METIS M311

  • Page 1 Pyrometers METIS M311 / M322 METIS H311 / H322 With 12-pin connector, LED or OLED display and push button device configuration Versions with integrated optics, optical fiber, PID controller, Profibus / Profinet Operating Manual...
  • Page 2 METIS M311 / M322 / H311 / H322 (12-pin) Content 1 General ..............4 5 Mechanical Installation ........16 1.1 Information to this Manual ....... 4 5.1 Mounting ............16 1.2 CE Conformity and Standards ......4 5.2 Mounting Optical Fiber / Optics (only devices with optical fiber) ..........
  • Page 3 METIS M311 / M322 / H311 / H322 (12-pin) 7.4 SensorTools Viewer: Graphical Measurement 6.2.13 Customize Analog Output (only via software SensorTools) ..........29 Data Playback..........53 6.2.14 Interface Type (only M3 devices) ....30 7.4.1 Load / Open Measurement Data File(s) ..53 6.2.15 Baud Rate...........
  • Page 4 Terms of Warranty A warranty period is 12 months from date of shipment form the Process Sensors facility. The seller will repair or replace the sensor at its own discretion. Further claims of the buyer against the seller or its agents are ex- cluded, especially claims for damages that are not incurred in the delivery itself.
  • Page 5 Customer Service / Spare Parts For technical information contact our customer service. It is recommended to purchase spare parts and accessories direct from Process Sensors Corporation. In addition, our employees are always interested in new information and experiences arising from the applica- tion and can be valuable for improvement of our products.
  • Page 6 METIS M311 / M322 / H311 / H322 (12-pin) Safety Use of the Operating Manual Any person who is tasked to carry out work with the device must have read and understood the operating in- structions before beginning. Operation and maintenance of the system may only be performed by trained per- sonnel.
  • Page 7 If the device is integrated into a system, the safety of the system in which the device is integrated is in the responsibility of the installer of the system. Process Sensors Corporation does not guarantee that the temperature control in all processes will meet the desired requirements. Process Sensors Corporation excludes the process responsibility.
  • Page 8 METIS M311 / M322 / H311 / H322 (12-pin) Overview Intended Use The here described 2-color pyrometers Metis M3 are devices for non-contact temperature measurement. With its short-wave spectral range they are suitable for measurements on metals, ceramics, graphite etc. with tem- perature ranges depending on the model execution between 300 and 3300°C (M3) or 350 and 3300°C (H3).
  • Page 9 METIS M311 / M322 / H311 / H322 (12-pin) Electrical Connection Cable Colors and Pin Assignment (14 wire connection cables) The electrical connection of the pyrometer (supply voltage and measuring signal) will be done via the connect- or on the rear panel. For this purpose, pre-assembled connection cables are available as accessories (see Accessories).
  • Page 10 METIS M311 / M322 / H311 / H322 (12-pin) 4.1.1 Factory Settings Parameters Factory Settings Display ..............2-color temperature Temperature sub range ..........Corresponds to basic range Response time t ............. Min (corresponds to: M3: <1 ms; H3: <80 µs) Emissivity slope ............
  • Page 11 METIS M311 / M322 / H311 / H322 (12-pin) Devices Devices with OLED with LED LED 4: Operating status:  Orange: self-test phase (M3 ca. 2 s; H3 during the initialization and thermostatic phase for ≤ 3 min.)  Green: Pyrometer is ready for operation. ...
  • Page 12 METIS M311 / M322 / H311 / H322 (12-pin) - Exceeding the maximum allowed internal device temperature - When equipped with PID controller: controller activity, control within definable setpoint limits, con- trol completed successfully, hold time finished.  Digital input: A voltage pulse of 24 V (exception: contin- M3 / H3 uous voltage at "Activate controller"), pulse length adjust- +24 V...
  • Page 13 METIS M311 / M322 / H311 / H322 (12-pin) Connecting several pyrometers via RS485: For a reflection-free operation with longer cables, pay attention to the correct cable termination. Termination at the physical bus is on front and rear. A (-) Master at B (+) the beginning:...
  • Page 14 METIS M311 / M322 / H311 / H322 (12-pin)  Open Device Manager (Windows 10: Right-click on the Windows key and select "Device Manager" in the menu. Windows 7: Open Control Panel and select Device Manager).  1. Double click on "Ports (COM and LPT)"  USB Serial Port (COMx) of the interface converter. ...
  • Page 15 METIS M311 / M322 / H311 / H322 (12-pin) Devices with additional Profibus-DP or Profinet connection The Profibus or Profinet interface is used for the digital communication of one or more pyrometers with a PLC. It is available in addition to the serial RS232 or RS485 interface. The connection is made in accordance with the applicable bus specifications.
  • Page 16 METIS M311 / M322 / H311 / H322 (12-pin) 4.2.1 Status LEDs for devices with Profibus / Profinet Profibus devices Profinet devices Not used Module status Operation mode Network status Modul status Link / activity LED for A Not used Link / activity LED for B Operation mode LED Meaning...
  • Page 17 METIS M311 / M322 / H311 / H322 (12-pin) |INFO| Through manufacturing tolerances the optical and mechanical axis are not running 100% in the same direction. The pyrometer should be realigned if it is installed again and twisted in the same holder (see Alignment onto the measuring object).
  • Page 18 METIS M311 / M322 / H311 / H322 (12-pin)  The colored heat shrink tube marker matches to the py- Protective tube rometer side.  The optics fit on the side without marking  Unscrew the light guide protective tube from the pyrom- eter’s housing Color mark Opt.
  • Page 19 METIS M311 / M322 / H311 / H322 (12-pin) Ratio Technology / Alignment onto the Measuring Object In opposition to radiation pyrometers, 2-color pyrometer measure in two spectral ranges simultaneously (two separate detectors at two adjacent narrow-band wavelengths) and determine the temperature by forming the radiation ratio (quotient).
  • Page 20 METIS M311 / M322 / H311 / H322 (12-pin) Turn on / off: Via the targeting light button on rear panel or via software Sensor- Tools (see 7.3.1, Control window). Laser target- (The targeting light will automatically switched off after 3 minutes if it is not ing button switched off manually.
  • Page 21 METIS M311 / M322 / H311 / H322 (12-pin) Setting the Measuring / Focus Distance The focus distance is the distance at which the lens has the smallest spot size. In most cases, this distance is also the required measuring distance. With focusable optics the focus distance can be changed continuously within a predetermined range, so the smallest possible spot size is always achieved.
  • Page 22 METIS M311 / M322 / H311 / H322 (12-pin) |NOTICE| Beware of tearing the optical fiber. Never Fixed optical fiber part pull on the flexible part of the optical fiber to focus! Flexible optical fiber part Spot Size Tables The following tables show the optical data of the different device types. The values in the tables are exemplary, intermediate measurement distances must be determined by interpolation.
  • Page 23 METIS M311 / M322 / H311 / H322 (12-pin) 5.6.2 Focusable Fiber-Optic Lenses Model M322: 300-1000°C M311/ M322: all Temp. ranges [°C] other temp. ranges H311: FSC <1500°C H311: FSC ≥1500°C H322: FSC <1200°C H322: FSC ≥1200°C 25 mm Optics Meas.
  • Page 24 METIS M311 / M322 / H311 / H322 (12-pin) Configuring the Pyrometer To configure the pyrometer (adapt to the measuring task and configure the connections), use the control but- tons on the back of the instrument When connected to a PC, all settings also can be made using the supplied SensorTools software (see 7, Sen- sorTools software) and, in addition, settings for the configurable inputs / outputs can be made;...
  • Page 25 METIS M311 / M322 / H311 / H322 (12-pin) Parameters / Settings The table shows the adjustable parameters in order they appear on the pyrometer, description next chapters. Parameters Meaning Settings MEASPARA.> Measurement parameters – Response time t MIN, 01 ms–10.0 s STMODE Storage mode peak picker: OFF / TIME / EXT.
  • Page 26 METIS M311 / M322 / H311 / H322 (12-pin) 6.2.1 Response Time t The response time specifies the time that the pyrometer requires to track 90% of a sudden temperature change on the signal output. The response time depends on the requirements of the respective measuring task and can be adjusted. ...
  • Page 27 METIS M311 / M322 / H311 / H322 (12-pin) curves are identical (at the same emissivity setting), then the emissivity slope is 1. Do they run parallel, so the correct value for the emissivity ratio must be found, such by comparative measurement with a thermocouple and subsequent ratio setting until the temperatures match.
  • Page 28 METIS M311 / M322 / H311 / H322 (12-pin) oriented measurements. They can Porcelain, glazed vary due to material-dependent condi- Porcelain, rough 0.8-0.9 0.8-0.9 tions, as in metals additionally to the Platinum 0.35 surface texture alloy components play Soot 0.95 0.95 an important role for the emissivity.
  • Page 29 METIS M311 / M322 / H311 / H322 (12-pin) 6.2.8 Temperature Sub Range The temperature range can be scaled down in order to adapt it to specific measurement conditions (minimum range 51°C), e.g. to configure the automatic clear function of the maximum value storage (see 6.2.2). A modi- fied temperature range also can be set for the analog outputs to increase the resolution even further.
  • Page 30 METIS M311 / M322 / H311 / H322 (12-pin)  Limit the analog output: if the analog output is set smaller than the pyrometer’s temperature range, the resolution of the analog output is increased even further.  Enlarge the analog output: Entering a larger range as the pyrometer’s range will scale the output. The out- put at the beginning of temperature range is 0 or4 ma, at the end of the pyrometer’s temperature range the current output is that corresponding to the enlarged temperature range.
  • Page 31 METIS M311 / M322 / H311 / H322 (12-pin) SensorTools Software  Installation With the minimum requirements devices can be connect and configured. When capturing or recording data, the performance is potentially impaired, that is errors or interruptions in data transfer can occur. With the recommended requirements all software features should be fully available.
  • Page 32 METIS M311 / M322 / H311 / H322 (12-pin) An identified pyrometer directly shows the current measuring temperature in the preselected measuring mode: 2C for 2-color temperature, in the control window changeable to C1 or C2 for 1-color temperature (channel 1 or channel 2).
  • Page 33 METIS M311 / M322 / H311 / H322 (12-pin)  The playback button starts the SensorTools viewer (see 7.4) and represents the last temporarily stored measurement data, since the device was connected. These are not stored on hard disk as in recording, only buffered.
  • Page 34 METIS M311 / M322 / H311 / H322 (12-pin)  Opens a control window for each connected device, if they have not been activated in the connect- ing window (devices overview tab). Scale per  Scale the graph: Beginning and end of temperature scale (left) and control output mouse click (right) scale can be adjusted to represent the relevant area reasonable Scale per mouse click: With the green arrows of the upper and lower...
  • Page 35 METIS M311 / M322 / H311 / H322 (12-pin) time, emissivity slope, emissivity, device temperature, setpoint if used, etc., and the higher the baud rate and transmission rate, the faster the value is reached. Data displayed in the SensorTools Viewer can be saved directly in the Viewer. ...
  • Page 36 METIS M311 / M322 / H311 / H322 (12-pin) |INFO| If an output is used, the assigned function can be used for controlling the automatic recording, for example for automatic recording when a certain temperature is exceeded (limit switch activated). Settings for the Fastest Possible Measurement Data Recording SensorTools can record very fast processes with recording intervals from approx.
  • Page 37 METIS M311 / M322 / H311 / H322 (12-pin) 7.3.1 Control Window  Measuring mode 2C means that the 2- Measured temperature Device temperature color temperature is displayed. In the Temp. measuring mode measured temperature selection field it can Set point temp. be changed to measuring temperature of (devices with PID Status of the digital...
  • Page 38 METIS M311 / M322 / H311 / H322 (12-pin) |INFO| Emissivity, transmittance and spot size fill factor are directly related to each other, the result of the multiplication of the 3 parameters has to be minimum 5%. If the multiplication of the 3 parameters falls be- low 5%, the values are displayed in red in the SensorTools control window.
  • Page 39 METIS M311 / M322 / H311 / H322 (12-pin) - Brightness: For manual brightness adjustment. - Brightness spot size: window size that should be used for the automatic brightness adjustment. Enter 5 to 100% of the window area, calculated centrally on the monitor center. |INFO| To prevent image over-modulation, the measured object should be in the middle (in the target mark) and the brightness spot size as large as the measuring object.
  • Page 40 METIS M311 / M322 / H311 / H322 (12-pin)  Device display enables to deactivate the display of the measuring temperature: Measurement temperature activated: shows the measured temperature on the display. Suppressed:  Devices with OLED display: display is switched off. ...
  • Page 41 METIS M311 / M322 / H311 / H322 (12-pin) - Targeting light on/off: (only available with equipment laser targeting light): Switches ex- Digital Input Digital Input Setups Active (x = function active) (x = input active) ternally on or off the laser targeting light. that can be Setup activated...
  • Page 42 METIS M311 / M322 / H311 / H322 (12-pin) Various settings: - Switching value: Activates the corresponding output at the set switching value (the LEDs 1-3 on the unit indicate the activated state). - Hysteresis: The hysteresis is the value at how many degree below the limit value the output is switched back.
  • Page 43 METIS M311 / M322 / H311 / H322 (12-pin) 7.3.2.2 Service functions To back-up, for printing or restore device parameters or create service files for remote diagnosis of problems.  Create configuration file: A configuration file ( data format * .r2p) contains all pyrometer data and device settings that have been made.
  • Page 44 METIS M311 / M322 / H311 / H322 (12-pin)  Display mode: - Single channel display: The control window displays the current temperature measurement, the device temperature and the setpoint value at devices equipped with a PID controller. - Multi-channel display: The SensorTools control window displays all temperatures (2-color, chan- nel 1, channel 2, device temperature) and the setpoint value at devices equipped with a PID...
  • Page 45 METIS M311 / M322 / H311 / H322 (12-pin) cur, the interval time can be increased, the baud rate can be increased or unnecessary programs and background services be stopped. A minor impact also has the reduction of the maximum data records in memory (see under 7.2.1 Main Window ...
  • Page 46 METIS M311 / M322 / H311 / H322 (12-pin) ten into the pyrometer. In this case the process should be performed again or the device should be reset to factory settings, this sets all calibration values back to factory calibration. ...
  • Page 47 METIS M311 / M322 / H311 / H322 (12-pin) 7.3.3 PID Controller Equipment M3 and H3 pyrometers can optionally be equipped with an integrated PID controller. At this, the temperature signal is converted into a control signal via a PID controller software. A setpoint value and the necessary PID parameters are set to trigger a corresponding control value output.
  • Page 48 METIS M311 / M322 / H311 / H322 (12-pin) - P-dynamic: Limits the power, the control output can control suddenly (standard 100%).  Other functions: - Setpoint limit 1 and 2: Activates every selected “setpoint limit” output (see 4.1.4 Configurable Inputs / Outputs and under 7.3.2 ...
  • Page 49 METIS M311 / M322 / H311 / H322 (12-pin) Fieldbus type (for info purposes) Fieldbus address (only with Profibus devices) Fieldbus mode selection Used and free storage Selected parameters (Sequence can be changed by moving) Transfer available parameters by moving them to the selec- tion list Save or load compiled lists on the PC...
  • Page 50 METIS M311 / M322 / H311 / H322 (12-pin) Mode 1 (Data packet size 64 Byte): Sequence and Meaning Byte Bytes Meaning Measured temperature (in 1/10°C); special case: F000h = invalid measured value (overflow) Emissivity (in ‰ ) 1000 = 100% Clear settings of maximum value storage / Storage mode (0...4) 0 = No storage;...
  • Page 51 METIS M311 / M322 / H311 / H322 (12-pin) Emissivity channel 1 (50…1200) Devices with controller: Proportional band Xp (0…10000) in 0.1% Emissivity channel 2 (50…1200) Devices with controller: Integral time Ti Emissivity sum channel (50…1200) (0…65535) Factor × sampling time Transmittance channel 1 (50…1000) Devices with controller: Derivative time Td Transmittance channel 2 (50…1000)
  • Page 52 METIS M311 / M322 / H311 / H322 (12-pin) 7.3.5 Setups Up to 7 different parameter settings (setups) can be stored in the pyrometer and activated via digital inputs (i.e. via external switch), (see 4.1.4 Configurable inputs / outputs). Depending on which and how many setups are to be used, one, two or all three inputs have to be used (see under 7.3.1 Control window ...
  • Page 53 METIS M311 / M322 / H311 / H322 (12-pin) SensorTools Viewer: Graphical Measurement Data Playback The SensorTools Viewer is an independent program for graphic display and evaluation of measured value da- ta. It can be opened by clicking on in the SensorTools graph tab or from the Windows pro- gram directory.
  • Page 54 METIS M311 / M322 / H311 / H322 (12-pin) 7.4.3 Customize Views / Measurement Data Evaluating Hide (inte- Open a separate information window for each meas- Show separate Draw selection grated) info info windows rectangle with the urement file and distribute it evenly and as large window as possible with the main window on the monitor individually...
  • Page 55 METIS M311 / M322 / H311 / H322 (12-pin) Communication / Options Prints a saved device configuration file ( data format *.r2p). To create the file, use the service function "Create configuration file" (see 7.3.2.2) while a pyrometer is connected. Update device Select program Search for software...
  • Page 56 METIS M311 / M322 / H311 / H322 (12-pin) SensorFlash Software (Update Firmware) The software SensorFlash is used to update the firmware of Process Sensors pyrometers. Press Firmware Update H3 / M3 in the Communication/Options tab (see 7.5) to start the software. Each SensorTools software has integrated the latest firmware files with the release date.
  • Page 57 METIS M311 / M322 / H311 / H322 (12-pin) Technical Data Device-specific M3 Model M311 M322 Temperature ranges 600–1400°C 900–2500°C 300–1000°C 600–2300°C 650–1500°C 1000–3000°C * 350–1300°C 800–3000°C ** 750–1800°C 1100–3300°C * 400–1600°C 1000–3300°C ** 500–1800°C Spectral range Channel 1: 0.93–1.1 µm / Channel 1: 1.65–1.8 µm / channel 2: 0.75–0.93 µm channel 2: 1.45–1.65 µm...
  • Page 58 METIS M311 / M322 / H311 / H322 (12-pin) Common Data Temp. sub ranges Any range adjustable within the temperature range (minimum span 50°C) Emissivity slope 0.800–1.200 Emissivity ε 0.050–1.200 (corresponds 5–120% in 0.1% steps) Transmission 0.050–1.000 (corresponds 5–100% in 0.1% steps) Spot size fill factor 0.050–1.000 (corresponds 5–100% in 0.1% steps) Switch-off level...
  • Page 59 METIS M311 / M322 / H311 / H322 (12-pin) Dimensions The case dimensions are the same for all models and differ in the mounting parts: Manual focusable optics 46.5 20.2 Motor focus optics Fiber optics devices, fiber optics 28.5 OQ25: optics 25 mm 26.5 17.5 21.5...
  • Page 60 METIS M311 / M322 / H311 / H322 (12-pin) Composition of the Type Number The equipment version of the pyrometer is encoded in the 22-digit model number as follows (Example type number M311 with 600–1300°C; laser targeting light, motorized focus, optics…): Digit 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 Expl.:...
  • Page 61 METIS M311 / M322 / H311 / H322 (12-pin) Accessories Ref. number Meaning AL11-XX Connection cable with right angled connector AL43-XX Connection cable with straight connector AL17- XX Connection cable with right angled connector + 0.5 m SC10 cable AM11-XX Connection cable with right angled connector + 1 m interface cable with SUB-D plug AM43-XX Connection cable with straight connector + 1 m interface cable with SUB-D plug...
  • Page 62 To obtain the measurement accuracy, we recommend to re-calibrate the pyrometer periodically (annually) with the help of calibration sources and optionally adjust at Process Sensors Corporation.  Fiber-optic pyrometers must always be calibrated together with the associated optics and the optical fiber cable.
  • Page 63 Communication via Serial Interface / Interface Commands Interface commands are used to communicate directly with a Process Sensors Corporation pyrometer. Com- mands can be used to write an own access control or can be entered in the terminal field of the software Sen- sorTools (see Communication / Options) or entered via a terminal software.
  • Page 64 METIS M311 / M322 / H311 / H322 (12-pin) Buffer mode 02: answer: AAAABBBBCCCCDDDDEEEEFFFFGGHHIIJJ AAAA = Temperature channel 1 (0xF001 => Overflow) BBBB = Temperature channel 2 (0xF001 => Overflow) CCCC = 2-color temperature (0xF001 => Overflow) DDDD = Setpoint value at ramp function (current ramp setpoint value!) EEEE = (Analog) control output (0-1000) in % (0.0-100.0%) FFFF = Signal strength (0-1000) in % (0.0-100.0%) GG = Data status byte 0...
  • Page 65 METIS M311 / M322 / H311 / H322 (12-pin) Storage mode (maximum value storage) 0 = No storage 1 = Time clearing maximum value storage 2 = External clearing of maximum value storage 3 = Automatic clearing of maximum value storage XXXXXX Clear time maximum value storage (in 100 µs steps) XXXXXX=0x000000...03d090 (corresponds 0...25 s)
  • Page 66 METIS M311 / M322 / H311 / H322 (12-pin) XYYYY Hysteresis for over temperature contact (uhX) (X=Digital output 1...3) to value YYYY (1/256°C) XYYYY Device over-temperature threshold at digital output X (ukX) X=1...3 to value YYYY (in 1/256°C) Read device identifier and date of software In addition, for devices with built-in PID controller: XXXX Hold time: XXXX=0x0000...ffff (in ms)
  • Page 67 METIS M311 / M322 / H311 / H322 (12-pin) Read camera status (answer: XX; 2-digit hex) Bit 0 = 1 Time appears Bit 6 = PAL Bit 12 = camera module active Bit 1 = 1 Date appears Bit 7 = NTSC Bit 13 = Unused Bit 2 = 1 Emissivity appears Bit 8 = Test image active...
  • Page 68 METIS M311 / M322 / H311 / H322 (12-pin) Index Maximum value storage / clear time t ...... 26 Measurement data playback ........53 Accessories ..............61 Mechanical installation ..........16 Address ............... 30 Model designs ............... 8 Alignment onto the measuring object ......19 Mounting ..............
  • Page 69 Manufacturer: Distributor: Sensortherm GmbH PROCESS SENSORS CORPORATION A KPM Analytics Company Hauptstraße 123 IR Temp. Sales: 787 Susquehanna Avenue, Franklin Lakes, NJ USA 65843 Sulzbach Tel / Fax: 201 485-8773 / -8770 Corporate Headquarters: 113 Cedar Street, Milford, MA USA Tel.