Endress+Hauser FXA291 Manual
Endress+Hauser FXA291 Manual

Endress+Hauser FXA291 Manual

Tof adapter
Hide thumbs Also See for FXA291:

Advertisement

BA01236F/00/EN/01.14
71240651
Products
ToF Adapter
FXA291
Solutions
Services

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FXA291 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Endress+Hauser FXA291

  • Page 1 Products Solutions Services BA01236F/00/EN/01.14 71240651 ToF Adapter FXA291...
  • Page 2: Table Of Contents

    7.5 Contact addresses for Endress+Hauser ........
  • Page 3: Safety Instructions

    Safety instructions Safety instructions Designated use The ToF Adapter is an adapter set for connecting the Commubox FXA291 to ToF devices, pressure transmitters and the Gammapilot M. The manufacturer does not accept liability for damages resulting from incorrect use or use other than that designated.
  • Page 4: Symbols Used

    Safety instructions ToF Adapter FXA291 Symbols used 1.5.1 Safety symbols Symbol Meaning DANGER! GEFAHR This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. A0011189-EN WARNING! WARNUNG This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
  • Page 5 ToF Adapter FXA291 Safety instructions 1.5.3 Symbols for types of information Symbol Meaning Allowed Indicates procedures, processes or actions that are allowed. A0011182 Preferred Indicates procedures, processes or actions that are preferred. A0011183 Forbidden Indicates procedures, processes or actions that are forbidden.
  • Page 6: Product Description

    Product description ToF Adapter FXA291 Product description Commubox FXA291 ToF Adapter FXA291 A0023045 Cable for device group 1 Cable for device group 2 Endress+Hauser...
  • Page 7: Incoming Acceptance And Product Identification

    ToF Adapter FXA291 Incoming acceptance and product identification Incoming acceptance and product identification Incoming acceptance Check the packaging and the contents for damage. Check that the goods supplied are complete and compare the scope of delivery with your order details.
  • Page 8: Scope Of Delivery

    EC declaration of conformity and thus complies with the statutory requirements of the EC directives. By applying the CE mark, Endress+Hauser confirms that the device has passed the necessary tests.
  • Page 9: Electrical Connection

    Electrical connection NOTICE Destruction of the device possible Use only cables pre-terminated by Endress+Hauser. The device is suitable only for connection to certified, intrinsically safe equipment with Ex ia type of protection. Connection data Intrinsically safe power circuit (e.g. Commubox FXA291) ...
  • Page 10: Connection With Devices Of Devices Group 2

    Electrical connection ToF Adapter FXA291 Connection with devices of devices group 2 4.3.1 Connecting to ToF device Connect the ToF Adapter FXA291 to the device' s display connector. Gammapilot M FMG60 Commubox FXA291 Levelflex M FMP4x Micropilot M FMR2xx Micropilot S FMR53x, FMR540...
  • Page 11: Operation

    ToF Adapter FXA291 Commubox FXA291 USB connection Operation is performed using a Personal Computer, a Commubox FXA291 with ToF Adapter FXA291 (USB) and the operating software "FieldCare". FieldCare is a graphical operating program for measuring devices from Endress+Hauser. It assists with commissioning, data back-up, signal analysis and documentation of the measuring point.
  • Page 12 ToF Adapter FXA291 FieldCare supports the following functions: • Configuration of transmitters in online operation • Loading and saving device data (upload/download) • Documentation of the measuring point Connection options: • Commubox FXA291 with ToF Adapter FXA291 via service interface Endress+Hauser...
  • Page 13: Commissioning

    ToF Adapter FXA291 Commissioning Commissioning NOTICE If using the ToF Adapter FXA291 with FDT/DTM software (e.g. FieldCare), the special IPC devices DTMs (according to the communication protocol) and „Level Pressure FXA193 CommDTM“ must be used. The DTMs are supplied on the FieldCare-CD-ROM.
  • Page 14: Technical Data

    Technical data ToF Adapter FXA291 Technical data Permitted ambient temperature range -20 °C ≤ T ≤ 70 °C Degree of protection IP30 Supplementary Documentation This documentation can be found on our product pages at "www.endress.com". 8.3.1 Certificates and approvals Ex approval The Ex certificates available can be found in the ordering information.
  • Page 15: Index

    ToF Adapter FXA291 Index CE mark ....... . . 8 Declaration of Conformity .
  • Page 16 71240651 www.addresses.endress.com...

Table of Contents