E2658-ETO 
Operating conditions 
The device should be used in a basic electromagnetic environment, where the latter is
 
 
defined in EN 61326-1. Avoid strong mechanical shock and vibrations. Although E2658
   
 
series devices are suitable for use in the presence of flammable dust, special precautions
 
 
should be taken when operating in a dusty environment. Dust may block the porous filter of
 
 
 
the sensor and compromise the results of the measurement. The detector should be
 
 
 
inspected regularly if used in a dusty atmosphere. Keep the instrument away from direct
 
sunlight. A sudden temperature or humidity change might affect the sensitivity of the
   
sensor. Make sure that installation and maintenance are performed in an explosion safe
 
 
atmosphere. 
Installation guidelines 
Before proceeding with the installation it is mandatory to read the Safety requirements
 
section and make sure to comply with all listed instructions. Installation standards EN
 
 
60079-29-2 and EN 45544-4 are also recommended for further instructions and related
 
 
information about the installation. During the installation of the detector-transmitter the
 
following points must be considered: 
Application (air quality control or leakage detection.) 
Properties of the space under investigation (room geometry, direction, and
velocity of airflows, etc.), 
Ethylene oxide is 1.5 times heavier than air. It is recommended to locate the
sensor not higher than the potential leakage. 
The device should be accessible for maintenance and repair. 
The aforementioned conditions above will affect the coverage area of the device. however,
 
the coverage area for a detector-transmitter is usually between 2.5 to 5 meters radius. 
For early leakage detection install the sensor as close as possible to the potential leakage
 
 
 
sources (flanges, valves, pressure reducers, pumps, etc.), taking into consideration other
 
points listed above.  
For general area monitoring without definite leakage sources, the detectors should be
 
 
 
distributed evenly in the room.  
For personal safety control, the detectors are installed in the breathing zone (at the height
 
 
of the head of people or animals). The recommended sensor position is vertical, pointing
 
 
   
downwards.  
 
 
     
 
 
 
 
 
   
 
     
 
 
   
     
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
     
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
     
 
 
   
 
 
   
 
 
 
 
 
 
   
 
 
 