Sensor probe handling  
The sensor probes of all types are equipped with a hydrophobic microporous PTFE filter to
 
 
protect the sensor from dust, dirt, and water drops. The filter should be replaced if it gets
 
 
strongly contaminated. To replace the PTFE filter, unscrew the M25 nut and remove the old
 
filter. Place a new filter into the nut and tighten it again. 
NOTE! Never stab or press the filter near its center where the sensor is located since this
 
 
may damage the sensor. Do not remove the filter as it may cause the device to show
 
 
incorrect values. 
The recommended orientation of the sensor probe is vertical with the sensor tip pointing
 
downwards. This prevents the possible accumulation of condensed water on the sensor
 
 
protection filter.  
Gas sensor replacement procedures 
1.
Remove the sensor head cap from the device, 
2.
Remove the PTFE filter (if it is not removed within the cap), 
3.
Remove the O-ring rubber, 
4.
Detach the electrochemical sensor from the device, 
5.
Insert the new electrochemical sensor inside, 
6.
Put back the O-ring rubber, PTFE filter, and the head cap, respectively. 
   
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
   
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 