Model 8901B
Installation
Steps
for
Installation.
Follow the procedure below to reinstall the Operating Information pull-out
card assembly:
1.
Install the information card by bowing it slightly in the middle and carefully guiding the edges
2. Push the information card
all the way into.the tray.
3. Place the information tray assembly between the rear feet of the instrument and slide it forward
4.
Replace the front feet of the instrument.
into the plastic guide slots near the front of the tray.
until the tabs are locked under the rear feet.
2-4. STORAGE AND SHIPMENT
Environment
The instrument should be stored in a clean, dry environment. The following environmental limitations
apply to both storage and shipment:
Temperature..
.........................................................
-55°C to +75OC
Humidity
................................................................
~ 9 5 % relative
Altitude..
...................................................
.15 300 meters (50 000 feet)
Packaging
Original Packaging.
Containers and materials identical to those used in factory packaging are available
through Hewlett-Packard offices. If the instrument is being returned to Hewlett-Packard for servicing,
attach a tag indicating the iype of service required, return address, model number and full serial
number. Also mark the container FRAGILE to assure careful handling. In any correspondence refer to
the instrument by model number and full serial number.
Other Packaging.
The following general instructions should be used for re-packaging with commer-
cially available materials:
1. Wrap the instrument in heavy paper or plastic.
(If shipping to a Hewlett-Packard office or service
center, attach a tag indicating the service required, return address, model number and
f u l l
serial
number.)
2. Use a strong shipping container.
A
double-wall carton made of 1.9 MPa (275 psi) test material
is adequate.
3. Use enough shock-absorbing material (75 to 100 mm layer; 3 to
4
in.) around all sides of
instrument to provide firm cushion and prevent movement in the container. Protect the front panel
with cardboard.
4.
Seal the shipping container securely.
5. Mark the shipping container FRAGILE to assure careful handling.
rev. 16JUN87
2-9