Appendix B Calibration
Factors
6-3
Display Code
Display Frequency
3 7 . 6 SPCL
Blue key,
tkHz
Verification
Procedure
Display Cal
Factor
%
Blue key,
MHz
Verify Reference
Cal Factor
Verify Frequency/Cal
Factor Pairs
1.
Press the green
AUTOMATIC OPERATION
key and then the
RF POWER
key. If the
HP
8901B displays Error 15, there is no
calibration data stored in the instrument; perform the Entry Procedure.
2.
Key
i n
37.5 SPCL,
the
Blue Shift
key, and then the
MHr
key. If the
displayed Reference Cal Factor does not match the Reference Cal Factor
(REF CAL FAC) on your sensor module, perform step 1 and step 6 in the
Entry Procedure.
Error Correction
Procedures
Cal Factor
0
If the frequency is correct but the cal factor is not, perform step 7
in the Entry Procedure for that frequency and enter its correct cal
factor only. (The HP 8901B
will
place the new entry in the correct
position in the table and remove the incorrect cal factor.) (For
example, if an incorrect cal factor value was entered for 10 MHz
you would correct it by entering
37.3 SPCL,
10
MHz,
the correct
cal factor value,
Blue Shift
key,
MHz.
The HP 8901B automatically
positions the new cal factor in the 10 MHz position in the table.)
Frequency
0
If an incorrect frequency is stored in the table, remove the
incorrect frequency by performing step 7 of the entry procedure
using the incorrect frequency and entering 0 as the cal factor.
Both
0
If you need to add both a frequency and its cal factor to the storage table,
perform the entry procedure (step
7)
for the correct frequency and cal factor.
(The HP 8901B
will
place the new entry in the correct position in the table.)